Effects of sterilization on the Tekscan digital pressure sensor.

Howard J Agins1, Valerie S Harder, Eugene P Lautenschlager, James C Kudrna.   

Abstract

Investigations into the effects of sterilization on a new biomechanical pressure sensor are necessary before contemplating in vivo use. Ten, designated Experimental, "K-Scan" digital pressure sensor arrays were sterilized with ethylene oxide gas (EtO), and their ability to accurately and reproducibly measure an applied load of 2225 N (500 lb) was assessed. Simultaneously, 10 un-sterilized sensor arrays, designated Control, were assessed. Each array was loaded 10 times inside a two-dimensional curved surface, and all arrays exhibited high reproducibility (coefficients of variation<2.0%). Following sterilization, the Experimental sensors showed a 22.2% average decrease in recorded force, a statistically significant difference from the pre-sterile data (p<0.002). However, when the Experimental sensors were re-calibrated post-sterilization, they showed only a 0.1% average decrease in recorded force, not a statistically significant difference (p>0.05, beta<0.05). Following 1-week storage, trial 2 data of the Control sensors showed a less dramatic yet significant 3.4% average decrease in recorded force when compared to trial 1 data (p<0.02). Control trial 2, once re-calibrated, showed a 0.5% average decrease in recorded force, not a statistically significant difference (p>0.05, beta<0.05). Results suggest that, following EtO sterilization, accurate and reproducible pressure measurements can be obtained from K-Scan sensors when calibration is performed at time of use.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14519350     DOI: 10.1016/s1350-4533(03)00119-x

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  2 in total

1.  Use of Tekscan K-scan sensors for retropatellar pressure measurement avoiding errors during implantation and the effects of shear forces on the measurement precision.

Authors:  A Wilharm; Ch Hurschler; T Dermitas; M Bohnsack
Journal:  Biomed Res Int       Date:  2013-12-03       Impact factor: 3.411

2.  Effect of simultaneous proprioceptive-visual feedback on gait of children with spastic diplegic cerebral palsy.

Authors:  Zeinab A Hussein; Ibrahim A Salem; Mostafa S Ali
Journal:  J Musculoskelet Neuronal Interact       Date:  2019-12-01       Impact factor: 2.041

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.