Abstract
The objective of this work was to determine if systematic differences exist between blood pressure time series measured by two non-invasive techniques. Cerebral blood flow autoregulation is often measured while a change in blood pressure is induced by deflation of thigh cuffs. To interpret the result a continuous measurement of arterial blood pressure is required. The Finapres is a non-invasive blood pressure monitor that is often used when assessing autoregulation but there is uncertainty about its reliability. A more recent device, the Colin tonometer, uses radial artery tonometry, which may prove to be a more reliable non-invasive method of obtaining a blood pressure waveform. Twenty healthy volunteers were recruited; blood pressure trends following cuff deflation were measured simultaneously with a Finapres and a Colin tonometer. A significant difference was found between the blood pressure waveforms measured with the two devices. The most striking difference was a slower recovery time measured with the Finapres (Wilcoxon signed rank test P < 0.001). Peripheral vasoaction may be distorting the measurement of blood pressure by the Finapres. This would account for the difference that exists between the techniques. Comparison with a direct arterial line would confirm which non-invasive measurement is more accurate.Mesh:
Year: 2003 PMID: 14509303 DOI: 10.1088/0967-3334/24/3/302
Source DB: PubMed Journal: Physiol Meas ISSN: 0967-3334 Impact factor: 2.833