K Kanemitsu1, H Kunishima, T Imasaka, S Ishikawa, H Harigae, S Yamato, Y Hirayama, M Kaku.
Abstract
We evaluated a low-temperature steam and formaldehyde (LTSF) sterilizer based on the draft European Standard prEN 14180. Microbiological tests were conducted on small and full loads using process challenge devices in five programs (P1-P5). With small loads all tests showed no growth of Bacillus stearothermophilus (ATCC7953) spores. However, positive cultures were observed with full-load tests using P5 (sterilization temperature, 50 degrees C). Our data indicated that the load influenced the efficacy of the LTSF sterilizer. Desorption tests were conducted to determine residual formaldehyde in indicator strips. The mean concentrations of formaldehyde in P1-P5 were 31.9, 56.3, 54.9, 82.2 and 180.6 microg, respectively, which are below the limits allowed by the draft Standard. Our results indicate that the LTSF sterilizer is useful for sterilization because of its excellent efficacy, short handling time, and safety.Entities:
Mesh:
Substances:
Year: 2003 PMID: 14505609 DOI: 10.1016/s0195-6701(03)00189-0
Source DB: PubMed Journal: J Hosp Infect ISSN: 0195-6701 Impact factor: 3.926