Literature DB >> 1448239

Rapid quality control of 99Tcm-sestamibi.

R M Reilly1, M So, J Polihronis, S Houle.   

Abstract

The thin layer chromatographic (TLC) method for determining the radiochemical purity of 99Tcm-sestamibi suggested by the manufacturer of the kit is slow and inconvenient. In this study a more rapid reversed phase chromatography (Sep-Pak) technique is compared with the TLC method for measurement of the radiochemical purity of 99Tcm-sestamibi. The levels of free 99Tcm-pertechnetate in 99Tcm-sestamibi were accurately determined using the Sep-Pak technique. However, due to binding of variable amounts of reduced-hydrolysed unbound 99Tcm to the chromatography cartridge, the Sep-Pak method overestimated the radiochemical purity of the radiopharmaceutical by approximately 3%. In practice, this overestimation is not important and the Sep-Pak technique can be used as a rapid method of determining radiochemical purity of 99Tcm-sestamibi.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1448239     DOI: 10.1097/00006231-199209000-00005

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  2 in total

1.  A miniaturized rapid paper chromatographic procedure for quality control of technetium-99m sestamibi.

Authors:  M Patel; S Sadek; S Jahan; A Owunwanne
Journal:  Eur J Nucl Med       Date:  1995-12

2.  Alternative chromatographic system for the quality control of lipophilic technetium-99m radiopharmaceuticals such as [(99m)Tc(MIBI)₆].

Authors:  D P Faria; C A Buchpiguel; F L N Marques
Journal:  Braz J Med Biol Res       Date:  2015-03-03       Impact factor: 2.590

  2 in total