Literature DB >> 1441694

[Peripheral embolism of hemostasis collagen (VasoSeal)].

G M Stiel1, C Beythien, H Kalkowski, K Hamper, H D Rohwer, C A Nienaber, C W Hamm.   

Abstract

VasoSeal is a purified bovine, absorbable collagen plug currently successfully used to close the femoral arterial puncture site after cardiac catheterization under full anticoagulation. Up to now there has been no experience with potential complications. We observed acute ischemia in the right lower leg of 2/100 patients 36 resp. 24 h after successful closure of the puncture site with VasoSeal. Angiography confirmed acute occlusion of the distal A. poplitea dextra. A 25-mm resp. 50-mm long cylindrical foreign body embolus was removed with a Fogarty-catheter by retrograde indirect embolectomy. Histopathology confirmed a fresh collagen clot with appositional thrombosis.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1441694

Source DB:  PubMed          Journal:  Z Kardiol        ISSN: 0300-5860


  2 in total

1.  Evaluation of the method of hemostasis after femoral arterial puncture.

Authors:  Jae-Hwan Cho; Hae-Kag Lee
Journal:  J Phys Ther Sci       Date:  2014-07-30

2.  Successful Retrieval of an Embolized Vascular Closure Device (Angio-Seal®) After Peripheral Angioplasty.

Authors:  Philipp Jud; Rupert Portugaller; Dennis Bohlsen; Thomas Gary; Marianne Brodmann; Gerald Hackl; Franz Hafner
Journal:  Cardiovasc Intervent Radiol       Date:  2017-01-18       Impact factor: 2.740

  2 in total