| Literature DB >> 1437332 |
S Arnon1, J Grigg, K Nikander, M Silverman.
Abstract
We compared the delivery of a micronized suspension of budesonide by a metered dose inhaler (MDI) with two different spacers (Aerochamber and Aerovent) and by two jet nebulizers (MAD2 and Ultravent) to a ventilated neonatal test-lung using a standard neonatal ventilator circuit. The combination of MDI and Aerochamber was significantly better at delivering budesonide to a filter in front of the test lung (14.2% of aerosolized dose) than were either the MDI and Aerovent (3.6%) or the Ultravent or MAD2 jet nebulizers (0.02% and 0.68% of initial reservoir dose). Of the droplets emerging from the MDI, Aerochamber, and ET tube, 18% of the initial dose was in droplets less than 4.7 microns. Assuming that the test-lung model accurately reflects in vivo deposition, the combination of MDI and Aerochamber appears to be an extremely effective way of delivering budesonide aerosol to ventilated newborn infants.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1437332 DOI: 10.1002/ppul.1950130309
Source DB: PubMed Journal: Pediatr Pulmonol ISSN: 1099-0496