A comparison of accuracy of computer treatment planning systems in brachytherapy.

P Feroldi1, M Galelli, S Belletti.   

Abstract

An intercomparison was made between dose-rate point calculations with five commercial computer treatment planning systems used in brachytherapy. Sixteen Italian radiotherapy medical physics institutions performed a twofold test, respectively for point and linear sources, reproducing actual clinical implants and the results were compared to references values. The test was designed in order to determine the errors introduced separately by computation and by implant reconstruction via orthogonal films. The results show that for point sources a better accuracy can be achieved than for linear sources. It is also shown that a large variation of results exists between computation systems and the variability is larger for linear sources. The digitizing procedure adds a large amount of error. In the whole set of calculation points, the overall percent difference between computed and reference dose-rate values is larger than +/- 5% in 18% cases for point and in 52% cases for linear sources, with data input by digitizer. It seems that relevant errors do occur when computing actual dose rates, resulting in clinically relevant inaccuracies in the calculated absorbed dose and in its relative effectiveness.

Mesh:

Year:  1992        PMID: 1410568     DOI: 10.1016/0167-8140(92)90373-3

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  1 in total

1.  Dosimetry and treatment planning of Occu-Prosta I-125 seeds for intraocular lesions.

Authors:  Suresh Chaudhari; Sudesh Deshpande; Vivek Anand; Sandeep De; Sanjay Saxena; A Dash; Mahua Basu; Preetam Samant; V Kannan
Journal:  J Med Phys       Date:  2008-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.