H V Allen, M F Anderson, J D Meindl.
Abstract
A totally implantable pulsed Doppler ultrasonic blood flowmeter has recently been developed to provide information on the velocity-flow profile in a vessel as well as its diameter. Volume flow can be indirectly obtained according to the formula: Q = (phi/4)(diam)2-v. In order to determine the accuracy of this estimate, in vivo direct bleedout measurements were performed on the abdominal aorta of six dogs with an overall accuracy in 77 trials of +2.0 +/- 8.7% (mean +/- 1 SD). The best-fit mean linear regression line was found to be: flowmeter output = 1.013-true flow + 5.1 ml/min. The scatter in the flowmeter's accuracy is thought to be due to small variations in the angle of the transducer. The source of the 2% overestimation in the mean accuracy could not be directly attributed to any one condition; the error is sufficiently small that in many cases it can be neglected.Entities:
Mesh:
Year: 1977 PMID: 140611 DOI: 10.1152/ajpheart.1977.232.5.H537
Source DB: PubMed Journal: Am J Physiol ISSN: 0002-9513