A new system for ambulatory pulmonary artery pressure recording.

J S Gibbs1, D MacLachlan, K M Fox.   

Abstract

OBJECTIVE: To develop a complete system for the measurement, recording, and analysis of ambulatory pulmonary artery pressure.
DESIGN: The new system consists of a pulmonary artery catheter, an ambulatory recorder, and a desktop computer. Pulmonary artery pressure is measured by a micromanometer tipped catheter with an in vivo calibration system to allow correction for zero drift. This catheter is plugged into a small battery powered recorder. The recorder has two input channels, one for pressure and one for an event marker. The pressure wave is sampled 32 times/s, processed by an in built computer, compressed, and stored in semi-conductor memory. On completion of a recording, data is transferred from the ambulatory recorder through a serial data link to an Acorn Archimedes desktop computer on which further data processing, statistical analysis, graphics, and printouts can be obtained.
RESULTS: The system has been used in 18 patients, with technically successful recording in 14, less than 15 minutes of data loss in three, and 12 hours of data loss in one.
CONCLUSIONS: A new system for ambulatory pulmonary artery monitoring has been developed and used clinically with success. It may provide new perspectives on the pathophysiology of disease as it applies to everyday life.

Entities:  

Mesh:

Year:  1992        PMID: 1389746      PMCID: PMC1025023          DOI: 10.1136/hrt.68.8.230

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  22 in total

1.  Evaluation of pulmonary arterial end-diastolic pressure as an indirect estimate of left atrial mean pressure.

Authors:  B S Jenkins; R D Bradley; M A Branthwaite
Journal:  Circulation       Date:  1970-07       Impact factor: 29.690

2.  Relationship of the pulmonary artery end-diastolic pressure to the left ventricular end-diastolic and mean filling pressures in patients with and without left ventricular dysfunction.

Authors:  R E Falicov; L Resnekov
Journal:  Circulation       Date:  1970-07       Impact factor: 29.690

3.  Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter.

Authors:  H J Swan; W Ganz; J Forrester; H Marcus; G Diamond; D Chonette
Journal:  N Engl J Med       Date:  1970-08-27       Impact factor: 91.245

4.  Continuous ambulatory pulmonary artery pressure monitoring. A new method using a transducer tipped catheter and a simple recording system.

Authors:  R D Levy; D Cunningham; L M Shapiro; C Wright; L Mockus; K M Fox
Journal:  Br Heart J       Date:  1986-04

5.  Continuous recording of pulmonary artery pressure in unrestricted subjects.

Authors:  H Ikram; A M Richards; E J Hamilton; M G Nicholls
Journal:  Br Heart J       Date:  1984-04

6.  Ambulatory monitoring of pulmonary artery pressure. A preliminary clinical evaluation.

Authors:  A W Nathan; S G Perry; T Cochrane; S O Banim; R A Spurrell; A J Camm
Journal:  Br Heart J       Date:  1983-01

7.  Syndrome X: the haemodynamic significance of ST segment depression.

Authors:  R D Levy; L M Shapiro; C Wright; L Mockus; K M Fox
Journal:  Br Heart J       Date:  1986-10

8.  Haemodynamic response to myocardial ischaemia during unrestricted activity, exercise testing, and atrial pacing assessed by ambulatory pulmonary artery pressure monitoring.

Authors:  R D Levy; L M Shapiro; C Wright; L Mockus; K M Fox
Journal:  Br Heart J       Date:  1986-07

9.  Diurnal variation in left ventricular function: a study of patients with myocardial ischaemia, syndrome X, and of normal controls.

Authors:  R D Levy; D Cunningham; L M Shapiro; C Wright; L Mockus; K M Fox
Journal:  Br Heart J       Date:  1987-02

10.  The haemodynamic significance of asymptomatic ST segment depression assessed by ambulatory pulmonary artery pressure monitoring.

Authors:  R D Levy; L M Shapiro; C Wright; L J Mockus; K M Fox
Journal:  Br Heart J       Date:  1986-12
View more
  4 in total

Review 1.  Implantable hemodynamic monitoring devices in heart failure.

Authors:  Salpy V Pamboukian; Melissa C Smallfield; Robert C Bourge
Journal:  Curr Cardiol Rep       Date:  2006-05       Impact factor: 2.931

2.  Pulmonary artery pressure variation in patients with connective tissue disease: 24 hour ambulatory pulmonary artery pressure monitoring.

Authors:  D A Raeside; G Chalmers; J Clelland; R Madhok; A J Peacock
Journal:  Thorax       Date:  1998-10       Impact factor: 9.139

3.  Ambulatory pulmonary artery pressure monitoring during sleep and exercise in normal individuals and patients with COPD.

Authors:  D A Raeside; A Brown; K R Patel; D Welsh; A J Peacock
Journal:  Thorax       Date:  2002-12       Impact factor: 9.139

Review 4.  Invasive Devices and Sensors for Remote Care of Heart Failure Patients.

Authors:  Sumant P Radhoe; Jesse F Veenis; Jasper J Brugts
Journal:  Sensors (Basel)       Date:  2021-03-12       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.