Literature DB >> 1388825

Mechanical complications of the reservoirs and flushing devices in ventricular shunt systems.

J F Martinez-Lage1, M Poza, J A Esteban.   

Abstract

Complications of valve reservoirs or flushing devices are seldom reported as a cause of shunt malfunction. The authors document their experience in six cases of mechanical complications of shunting devices. Two cases presented with collapse and intracranial migration of the valve reservoir, a complication that has not been reported previously. In two cases, the plastic dome became disconnected from the metallic base of the reservoir. A fifth patient, with an integral shunt system, showed a fracture in the soldered join of the distal tube to the reservoir dome. The last patient had the valve reservoir partially collapsed by bone growth at the time of shunt revision.

Entities:  

Mesh:

Year:  1992        PMID: 1388825     DOI: 10.3109/02688699209023790

Source DB:  PubMed          Journal:  Br J Neurosurg        ISSN: 0268-8697            Impact factor:   1.596


  5 in total

1.  Cerebrospinal fluid shunt valve fracture with separation: report of three cases.

Authors:  L C Hellbusch
Journal:  Childs Nerv Syst       Date:  1996-03       Impact factor: 1.475

2.  Rationale and methodology of the multicenter pediatric cerebrospinal fluid shunt design trial. Pediatric Hydrocephalus Treatment Evaluation Group.

Authors:  J M Drake; J Kestle
Journal:  Childs Nerv Syst       Date:  1996-08       Impact factor: 1.475

3.  Ventriculoperitoneal Shunt Malfunction in a Pediatric Patient Due to Compression by Skull Growth: A Case Report.

Authors:  Masamichi Endo; Shunya Hanakita; Soichi Oya
Journal:  NMC Case Rep J       Date:  2022-04-21

4.  Subdural hemorrhage of the cauda equina. A rare complication of cerebrospinal fluid shunt. Case report.

Authors:  G Wurm; P Pogady; K Lungenschmid; J Fischer
Journal:  Neurosurg Rev       Date:  1996       Impact factor: 3.042

5.  Fractured inlet connecting tube of the flat bottom flushing device of a posterior fossa cystoperitoneal shunt.

Authors:  Abbas Amirjamshidi; Kazem Abbasioun; Babak Ghassemi
Journal:  Surg Neurol Int       Date:  2015-03-05
  5 in total