| Literature DB >> 1366447 |
J M Fernández-Romero1, M D Luque de Castro, M Valcárcel.
Abstract
Open-closed flow injection systems allow the development of methods based on multidetection by using conventional detectors. The methods proposed here for the determination of alanine aminotransferase (ALT) utilize a photometric detector included in the closed circuit, which allows one to perform fixed-time and reaction-rate measurements whose sensitivity can be increased by using the sum of the analytical signals. The methods, with determination ranges between 1 and 500 U l-1 of ALT, feature sampling frequencies up to 60 h-1. Their application to serum samples provided excellent results, with recoveries between 95 and 106% and averaging at 99.9%.Entities:
Mesh:
Substances:
Year: 1990 PMID: 1366447 DOI: 10.1016/0168-1656(90)90016-5
Source DB: PubMed Journal: J Biotechnol ISSN: 0168-1656 Impact factor: 3.307