Literature DB >> 1358481

Analysis for transaminases in serum with an amperometric glutamate electrode.

D Compagnone1, G Federici, R Massoud, L Santoro, M Anichini, G Palleschi.   

Abstract

We determined transaminases in human blood serum with an amperometric glutamate biosensor. The probe was a hydrogen peroxide sensor assembled with appropriate selective membranes to enhance the probe specificity and lifetime. Calibration curves of glutamate were linear in the range 1-1000 mumol/L, with a response time of < 1 min. This probe was subsequently applied to the measurement of activities of aspartate and alanine aminotransferases in human sera. Analytical recovery studies demonstrated the suitability of the glutamate sensor by measuring 91-99% of added glutamate, 92-106% of added aspartate aminotransferase, and 101-105% of added alanine aminotransferase. Transaminase activity measured in 80 sera correlated well with results obtained with a spectrophotometric procedure.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1358481

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  A Micro-Platinum Wire Biosensor for Fast and Selective Detection of Alanine Aminotransferase.

Authors:  Tran Nguyen Thanh Thuy; Tina T-C Tseng
Journal:  Sensors (Basel)       Date:  2016-05-26       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.