Evaluation of 2 scatter correction methods using a striatal phantom for quantitative brain SPECT.

Douglass C Vines1, Masanori Ichise, Jeih-San Liow, Hiroshi Toyama, Robert B Innis.   

Abstract

OBJECTIVE: Scatter correction is an important factor in quantitative SPECT. In this study, we evaluated 2 methods of scatter correction for brain SPECT. The first is based on thresholding the energy spectrum (ES), and the second is based on a modification of the transmission-dependent convolution subtraction (TDCS) method.
METHODS: SPECT imaging of a skull striatal phantom was performed using a triple-head camera with and without scatter correction. The striatal compartments were filled with (123)I, and the brain shell cavity (background) was filled with varying concentrations of (123)I to obtain striatal-to-background ratios of 2, 5, 10, 15, 20, and 25 to 1, respectively, which were considered to be the expected ratios. SPECT-measured ratios of striatal-to-background counts were determined with scatter correction (both ES and TDCS methods) and without scatter correction and were then compared with the expected ratios.
RESULTS: Without scatter correction, measured striatal-to-background ratios were underestimated by an average of 41.7%, compared with the expected ratios. The ES method of scatter correction underestimated the striatal-to-background ratios by an average of 27.4%, a significant improvement (P < 0.04) over those without scatter correction. With the TDCS method of scatter correction, the ratios were underestimated by only 3.3% (P < 0.03). TDCS ratios were significantly (P < 0.04) higher than ES ratios and were nearly identical to the expected ratios.
CONCLUSION: These results suggest that scatter correction significantly improves the striatal-to-background ratios. The TDCS method appears to correct scatter more effectively than does the ES method for the striatal phantom, thus providing more accurate quantification.

Entities:  

Mesh:

Year:  2003        PMID: 12968046

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  2 in total

1.  Individual voxelwise dosimetry of targeted 90Y-labelled substance P radiotherapy for malignant gliomas.

Authors:  Stefan Kneifel; Peter Bernhardt; Helena Uusijärvi; Stephan Good; Ludwig Plasswilm; Carlos Buitrago-Téllez; Jan Müller-Brand; Helmut Mäcke; Adrian Merlo
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-01-31       Impact factor: 9.236

2.  Practical issues and limitations of brain attenuation correction on a simultaneous PET-MR scanner.

Authors:  J E Mackewn; J Stirling; S Jeljeli; S-M Gould; R I Johnstone; I Merida; L C Pike; C J McGinnity; K Beck; O Howes; A Hammers; P K Marsden
Journal:  EJNMMI Phys       Date:  2020-05-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.