Dosimetric effect of respiration-gated beam on IMRT delivery.

Jun Duan1, Sui Shen, John B Fiveash, Ivan A Brezovich, Richard A Popple, Prem N Pareek.   

Abstract

Intensity modulated radiation therapy (IMRT) with a dynamic multileaf collimator (DMLC) requires synchronization of DMLC leaf motion with dose delivery. A delay in DMLC communication is known to cause leaf lag and lead to dosimetric errors. The errors may be exacerbated by gated operation. The purpose of this study was to investigate the effect of leaf lag on the accuracy of doses delivered in gated IMRT. We first determined the effective leaf delay time by measuring the dose in a stationary phantom delivered by wedge-shaped fields. The wedge fields were generated by a DMLC at various dose rates. The so determined delay varied from 88.3 to 90.5 ms. The dosimetric effect of this delay on gated IMRT was studied by delivering wedge-shaped and clinical IMRT fields to moving and stationary phantoms at dose rates ranging from 100 to 600 MU/min, with and without gating. Respiratory motion was simulated by a linear sinusoidal motion of the phantom. An ionization chamber and films were employed for absolute dose and 2-D dose distribution measurements. Discrepancies between gated and nongated delivery to the stationary phantom were observed in both absolute dose and 2-D dose distribution measurements. These discrepancies increased monotonically with dose rate and frequency of beam interruptions, and could reach 3.7% of the total dose delivered to a 0.6 cm3 ion chamber. Isodose lines could be shifted by as much as 3 mm. The results are consistent with the explanation that beam hold-offs in gated delivery allowed the lagging leaves to catch up with the delivered monitor units each time that the beam was interrupted. Low dose rates, slow leaf speeds and low frequencies of beam interruptions reduce the effect of this delay-and-catch-up cycle. For gated IMRT it is therefore important to find a good balance between the conflicting requirements of rapid dose delivery and delivery accuracy.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12945990     DOI: 10.1118/1.1592017

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Feasibility study on inverse four-dimensional dose reconstruction using the continuous dose-image of EPID.

Authors:  Inhwan Jason Yeo; Jae Won Jung; Byong Yong Yi; Jong Oh Kim
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

Review 2.  Advances in the use of motion management and image guidance in radiation therapy treatment for lung cancer.

Authors:  Jason K Molitoris; Tejan Diwanji; James W Snider; Sina Mossahebi; Santanu Samanta; Shahed N Badiyan; Charles B Simone; Pranshu Mohindra
Journal:  J Thorac Dis       Date:  2018-08       Impact factor: 2.895

3.  Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy.

Authors:  C Khamfongkhruea; C Tannanonta; S Thongsawad
Journal:  Biomed Imaging Interv J       Date:  2012-01-01

4.  Dosimetric characteristics of a new linear accelerator under gated operation.

Authors:  Sergey Kriminski; Alex N Li; Timothy D Solberg
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.