Ronald O Rahn1, Mihaela I Stefan, James R Bolton, Evan Goren, Ping-Shine Shaw, Keith R Lykke.
Abstract
The quantum yield (QY) of the iodide-iodate chemical actinometer (0.6 M KI-0.1 M KIO3) was determined for irradiation between 214 and 330 nm. The photoproduct, triiodide, was determined from the increase in absorbance at 352 nm, which together with a concomitant measurement of the UV fluence enabled the QY to be calculated. The QY at 254 nm was determined to be 0.73 +/- 0.02 when calibration was carried out against a National Institute of Standards and Technology traceable radiometer or photometric device. At wavelengths below 254 nm the QY increased slightly, leveling off at -0.80 +/- 0.05, whereas above 254 nm the QY decreases linearly with wavelength, reaching a value of 0.30 at 284 nm. In addition, the QY was measured at different iodide concentrations. There is a slight decrease in QY going from 0.6 to 0.15 M KI, whereas below 0.15 M KI the QY drops off sharply, decreasing to 0.23 by 0.006 M KI. Calibration of the QY was also done using potassium ferrioxalate actinometry to measure the irradiance. These results showed a 20% reduction in QY between 240 and 280 nm as compared with radiometry. This discrepancy suggests that the QY of the ferrioxalate actinometer in this region of the spectrum needs reexamination.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12945582 DOI: 10.1562/0031-8655(2003)078<0146:qyotic>2.0.co;2
Source DB: PubMed Journal: Photochem Photobiol ISSN: 0031-8655 Impact factor: 3.421