Literature DB >> 12945278

[The sensitized reaction of non-ionic microemulsion on the determination of trace Ni(II) by fluorescence quenching method with salicyfluorone].

X Zhu1, R Guo, X Zhang.   

Abstract

A new selective and sensitive spectrofluorimetric method in non-ionic microemulsion for the determination of trace Ni(II) in water has been developed. This method is based on the quenching of fluorescence of Triton X-100/n-C5H11OH/n-C9H20/H2O microemulsion system due to the formation of complex Ni(II)-SAF-Triton X-100. The results show that there is a linear relationship in the rang of 0.5-2.5 ng.25 mL-1 for Ni(II). The sensitivity in microemulsion is higher than that in micelle. Satisfactory results are obtained for Ni(II) in water sample.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 12945278

Source DB:  PubMed          Journal:  Guang Pu Xue Yu Guang Pu Fen Xi        ISSN: 1000-0593            Impact factor:   0.589


  1 in total

1.  Determination of iron(III) by room temperature ionic liquids/surfactant sensitized fluorescence quenching method.

Authors:  Xiashi Zhu; Rongrong Jiang
Journal:  J Fluoresc       Date:  2010-10-19       Impact factor: 2.217

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.