Literature DB >> 12941504

Comparison of methods for preparation of 125I brachytherapy source cores for the treatment of eye cancer.

R B Manolkar1, S U Sane, K T Pillai, M A Majali.   

Abstract

Two procedures for fixing the 125I activity on different matrices were explored. In the first method, anodic deposition of 125I on silver wire was used. In the second, 125I was adsorbed as iodate (IO3-) on alumina microspheres by using a solid-solution interface technique. While the 125I uptake was approximately 84% on the silver wires by electrodeposition method, the microspheres showed approximately 97% by the solid-solution interface technique. The low leachable (0.05%) wires, and comparatively high leachable (approximately 4%) spheres were trial encapsulated and laser welded. The encapsulated sources were measured for source strength and were found to be suitable for the treatment of eye cancer.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12941504     DOI: 10.1016/s0969-8043(03)00152-0

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Validation of American Association of Physicists in Medicine TG 43 Dosimetry Data in Commercial Treatment Planning System.

Authors:  Sathiyan Saminathan; K Maheshraja; K M Ganesh
Journal:  J Med Phys       Date:  2021-09-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.