Elastographic measurement of the area and volume of thermal lesions resulting from radiofrequency ablation: pathologic correlation.

Tomy Varghese1, Udomchai Techavipoo, Wu Liu, James A Zagzebski, Quan Chen, Gary Frank, Fred T Lee.   

Abstract

OBJECTIVE: Elastography is a promising tool for visualizing the zone of necrosis in liver tissue resulting from radiofrequency tumor ablation. Because heat-ablated tissues are stiffer than normal untreated tissue, elastography may prove useful for following up patients who undergo radiofrequency ablative therapy. We sought to report the initial evaluations of the reliability of elastography for delineating thermal lesion boundaries in liver tissue by comparing lesion dimensions determined by elastography with the findings at whole-mount pathology.
MATERIALS AND METHODS: Radiofrequency ablation was performed in vitro on liver tissue specimens encased in gelatin phantoms. The imaging plane for elastography was perpendicular to the axis of the radiofrequency electrode so that the ablated region was around the center of the plane. To obtain three-dimensional visualization of thermal lesions, we reconstructed the lesions from multiple elastograms by linearly translating the elastographic scanning plane. Pathology photographs were obtained in the same image plane used for elastography by slicing through the gelatin and tissue phantom using external markers. We used digitized gross pathology photographs obtained at a specified slice thickness to compute the areas and volumes of the lesions. These measurements were then compared to the measurements obtained from the elastograms.
RESULTS: In a sample of 40 thermal lesions, we obtained a correlation between in vitro elastographic and pathologic measurements of r = 0.9371 (p < 0.00001) for area estimates and r = 0.979 (p < 0.00001) for volume estimates.
CONCLUSION: We found excellent correlation between the measurements of the dimensions, areas, and volumes of thermal lesions that were based on elastographic images and the measurements that were based on digitized pathologic images. When compared with digitized pathologic measurements, elastographic measurements showed a tendency to slightly underestimate both the areas and volumes of lesions. Nevertheless, elastography is a reliable technique for delineating thermal lesions resulting from radiofrequency ablation.

Entities:  

Mesh:

Year:  2003        PMID: 12933463     DOI: 10.2214/ajr.181.3.1810701

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  29 in total

1.  Quantifying local stiffness variations in radiofrequency ablations with dynamic indentation.

Authors:  Ryan J DeWall; Tomy Varghese; Christopher L Brace
Journal:  IEEE Trans Biomed Eng       Date:  2011-12-08       Impact factor: 4.538

2.  Electrode displacement strain imaging of thermally-ablated liver tissue in an in vivo animal model.

Authors:  N Rubert; S Bharat; R J DeWall; A Andreano; C Brace; J Jiang; L Sampson; T Varghese
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

3.  Shear wave velocity imaging using transient electrode perturbation: phantom and ex vivo validation.

Authors:  Ryan J DeWall; Tomy Varghese; Ernest L Madsen
Journal:  IEEE Trans Med Imaging       Date:  2010-11-11       Impact factor: 10.048

4.  Improvements in elastographic contrast-to-noise ratio using spatial-angular compounding.

Authors:  Udomchai Techavipoo; Tomy Varghese
Journal:  Ultrasound Med Biol       Date:  2005-04       Impact factor: 2.998

5.  Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities.

Authors:  B J Fahey; S J Hsu; P D Wolf; R C Nelson; G E Trahey
Journal:  Phys Med Biol       Date:  2006-07-20       Impact factor: 3.609

6.  Challenges and implementation of radiation-force imaging with an intracardiac ultrasound transducer.

Authors:  Stephen J Hsu; Brian J Fahey; Douglas M Dumont; Patrick D Wolf; Gregg E Trahey
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2007-05       Impact factor: 2.725

7.  Segmentation of elastographic images using a coarse-to-fine active contour model.

Authors:  Wu Liu; James A Zagzebski; Tomy Varghese; Charles R Dyer; Udomchai Techavipoo; Timothy J Hall
Journal:  Ultrasound Med Biol       Date:  2006-03       Impact factor: 2.998

8.  Radio-frequency ablation electrode displacement elastography: a phantom study.

Authors:  Shyam Bharat; Tomy Varghese; Ernest L Madsen; James A Zagzebski
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

9.  Real-time sonoelastography of hepatic thermal lesions in a swine model.

Authors:  Man Zhang; Benjamin Castaneda; Jared Christensen; Wael E Saad; Kevin Bylund; Kenneth Hoyt; John G Strang; Deborah J Rubens; Kevin J Parker
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

10.  Contrast in intracardiac acoustic radiation force impulse images of radiofrequency ablation lesions.

Authors:  Stephanie A Eyerly; Tristram D Bahnson; Jason I Koontz; David P Bradway; Douglas M Dumont; Gregg E Trahey; Patrick D Wolf
Journal:  Ultrason Imaging       Date:  2014-04       Impact factor: 1.578

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.