Literature DB >> 12930497

Diagnostic yield of external electrocardiographic loop recorders in patients with recurrent syncope and negative tilt table test.

Andreas Schuchert1, Renke Maas, Christina Kretzschmar, Gina Behrens, Ingrid Kratzmann, Thomas Meinertz.   

Abstract

External electrocardiographic loop recording permits extended cardiac rhythm monitoring and ECG storage before and after device activation. The purpose of the study was to assess the diagnostic yield of external loop recorders in patients with more than two syncopal events within the last 6 months and a negative tilt table test. Twenty-four consecutive patients (51 +/- 14 years, male n = 9) were enrolled with 3 +/- 4 recurrent syncopal events within the last 6 months and no overt structural heart disease. The loop recorder CardioCall continuously records a two-channel ECG via skin electrodes. When activated, up to 4.5 minutes of the ECG preceding activation is stored. The time between tilt table testing and monitoring was 5 +/- 4 weeks. The average monitoring period covered 50 +/- 22 days. Recording was either unsuccessful or terminated prematurely in 5 patients. The reasons were skin irritation secondary to the electrodes in two patients, cable damage in one patient, and two patients inadvertently erased the stored data when replacing the batteries. Fourteen patients activated the loop recorder at least once. Symptoms leading to device activation were syncope (n = 1), dizziness (n = 11), feeling unwell (n = 7), palpitations (n = 5), headaches (n = 1), and chest pain (n = 1). The loop recorder documented sinus tachycardia during the syncopal event. Sinus tachycardia was also observed in 7 other patients, and atrial flutter in two patients. Eight patients had recurrent syncope with two patients experiencing syncope prior to the monitoring period, one syncope occurred in a patient who inadvertently erased the stored data, one event coincided with sinus tachycardia, and 4 (17%) patients had syncope during 15 +/- 10 months after termination of loop monitoring. The external loop recorder was not very useful for arrhythmia detection in patients with syncopal events, no overt heart disease, and a negative tilt table test because the cardiac rhythm was stored in only 1 of 8 (13%) patients with recurrent syncope. Reasons for the low diagnostic yield of external loop recorders were infrequent syncopal events after the baseline evaluation, with rare events during the limited monitoring period in particular, and premature termination or unsuccessful recording in 21% of patients.

Entities:  

Mesh:

Year:  2003        PMID: 12930497     DOI: 10.1046/j.1460-9592.2003.t01-1-00277.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  13 in total

Review 1.  How to avoid a misdiagnosis in patients presenting with transient loss of consciousness.

Authors:  Sanjiv Petkar; Paul Cooper; Adam P Fitzpatrick
Journal:  Postgrad Med J       Date:  2006-10       Impact factor: 2.401

2.  Guidelines for the diagnosis and management of syncope (version 2009).

Authors:  Angel Moya; Richard Sutton; Fabrizio Ammirati; Jean-Jacques Blanc; Michele Brignole; Johannes B Dahm; Jean-Claude Deharo; Jacek Gajek; Knut Gjesdal; Andrew Krahn; Martial Massin; Mauro Pepi; Thomas Pezawas; Ricardo Ruiz Granell; Francois Sarasin; Andrea Ungar; J Gert van Dijk; Edmond P Walma; Wouter Wieling
Journal:  Eur Heart J       Date:  2009-08-27       Impact factor: 29.983

3.  [A patient with syncope].

Authors:  Helene Höhler
Journal:  Med Klin (Munich)       Date:  2009-10-15

Review 4.  [Event-recorder].

Authors:  Andreas Schuchert
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2008-10-25

Review 5.  [Diagnostics of patients with syncope of unknown origin: indications for hospital admission and for in- and out-hospital services?].

Authors:  Andreas Schuchert; Ekhard Wille
Journal:  Herz       Date:  2008-09       Impact factor: 1.443

6.  Development of a diagnostic protocol for dizziness in elderly patients in general practice: a Delphi procedure.

Authors:  Otto R Maarsingh; Jacquelien Dros; Henk C van Weert; François G Schellevis; Patrick J Bindels; Henriette E van der Horst
Journal:  BMC Fam Pract       Date:  2009-02-07       Impact factor: 2.497

7.  Using the net benefit regression framework to construct cost-effectiveness acceptability curves: an example using data from a trial of external loop recorders versus Holter monitoring for ambulatory monitoring of "community acquired" syncope.

Authors:  Jeffrey S Hoch; Marie Antoinette Rockx; Andrew D Krahn
Journal:  BMC Health Serv Res       Date:  2006-06-06       Impact factor: 2.655

8.  External prolonged electrocardiogram monitoring in unexplained syncope and palpitations: results of the SYNARR-Flash study.

Authors:  E T Locati; A Moya; M Oliveira; H Tanner; R Willems; M Lunati; M Brignole
Journal:  Europace       Date:  2015-10-29       Impact factor: 5.214

9.  Cardiovascular assessment of falls in older people.

Authors:  Maw Pin Tan; Rose Anne Kenny
Journal:  Clin Interv Aging       Date:  2006       Impact factor: 4.458

10.  Clinical symptoms associated with asystolic or bradycardic responses on implantable loop recorder monitoring in patients with recurrent syncope.

Authors:  Khalil Kanjwal; Yousuf Kanjwal; Beverly Karabin; Blair P Grubb
Journal:  Int J Med Sci       Date:  2009-04-09       Impact factor: 3.738

View more