Literature DB >> 12914636

Lead extraction via the femoral artery of a left ventricular pacing lead inserted in the subclavian artery.

Frank Bracke1, Albert Meijer.   

Abstract

We describe the extraction of a pacing lead via the femoral artery 5 months after it was inadvertently introduced through the right subclavian artery and positioned in the left ventricle. The lead was dislocated from the myocardium by traction from the subpectoral area and subsequently removed with a retriever inserted through the femoral artery in order to minimize the risk of cerebral embolization and bleeding at the entry site. Bleeding was controlled at the subclavian artery entry site by temporary inflation of a balloon catheter inserted through the brachial artery, and at the insertion site of the retriever in the femoral artery with a closure device.

Entities:  

Mesh:

Year:  2003        PMID: 12914636     DOI: 10.1046/j.1460-9592.2003.t01-1-00225.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Pacing lead inserted via the subclavian artery caused acute coronary syndrome.

Authors:  Peter Nordbeck; Heiner Langenfeld; Axel Krein; Wolfgang R Bauer; Oliver Ritter
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-02-11

2.  Inadvertent transarterial lead placement in the left ventricle and aortic cusp: percutaneous lead removal with carotid embolic protection and stent graft placement.

Authors:  Ioanna Kosmidou; Dimitri Karmpaliotis; David E Kandzari; Dan Dan
Journal:  Indian Pacing Electrophysiol J       Date:  2012-12-02
  2 in total