Mechanical properties of various z-stent designs: an endovascular stent-grafting perspective.

Abul Hasan Muhammad Bashar1, Teruhisa Kazui, Naoki Washiyama, Hitoshi Terada, Katsushi Yamashita, Mohammad E Haque.   

Abstract

PURPOSE: To comparatively assess the mechani-cal behavior of various clinically relevant Z-stent designs.
METHODS: A total of 16 Z-stents of original, biliary, spiral, and double-skirted designs (n=4 for each) were constructed using similar specifications for all. Stents were then evaluated for stiffness, snap opening force (SOF), flexibility, and displacement force using a novel tensiometer. Differences among the stents were determined using statistical methods. Stents explanted from dog aorta after a mean follow-up of 13 months were examined under a scanning electron microscope for surface defects.
RESULTS: Forces required for about 50% reduction in diameter were 1.88 +/- 0.16 N, 3.81 +/- 0.21 N, 2.76 +/- 0.22 N, and 3.35 +/- 0.19 N for original, biliary, spiral, and skirted designs, respectively. Differences among the four designs were statistically significant at almost all points of measurement (P < 0.0001). Biliary and skirted designs showed higher SOF values in the early measurements. Stents explanted from dog aorta after a mean duration of 13 months showed no obvious corrosion or breakage in the wire struts.
CONCLUSIONS: Significant differences exist among the various Z-stent designs in terms of their mechanical properties. Understanding them should help to select the appropriate stent for a given lesion. 316L stainless steel shows a favorable long-term tissue interaction.

Entities:  

Mesh:

Year:  2003        PMID: 12911346     DOI: 10.1046/j.1525-1594.2003.06995.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094