| Literature DB >> 12907263 |
Ming Yao1, Hongjian Zhang, Saeho Chong, Mingshe Zhu, Richard A Morrison.
Abstract
Digoxin is a cardiac glycoside that is widely used for the treatment of congestive heart failure. To evaluate pharmacokinetics of digoxin in rats, a sensitive LC/MS/MS assay was developed and validated for the determination of digoxin concentration in rat plasma. For detection, a Sciex API3000 LC/MS/MS with atmospheric pressure ionization (API) mass spectrometry turbo ion spray inlet in the positive ion-multiple reaction monitoring mode was used to monitor precursor-->product ions of m/z 798.6-->651.6 for digoxin and m/z 577.6-->433.3 for oleandrin, the internal standard (IS). The standard curve was linear (r(2)>or=0.999) over the digoxin concentration range of 0.1-100 ng/ml in plasma for digoxin. The mean predicted concentrations of the quality control samples deviated by <5.8% from the corresponding nominal values; the intra-assay and inter-assay precision of the assay were within 8.6% relative standard deviation. At the lower limit of quantitation (LLQ) of 0.1 ng/ml, the mean deviation of predicted concentrations from the nominal value was within 3.7%. The extraction recoveries of digoxin and internal standard were 82.7+/-3.9 and 105.9+/-2.3%, respectively. The present method was successfully applied to characterization of pharmacokinetic profiles of digoxin in rats after oral administration.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12907263 DOI: 10.1016/s0731-7085(03)00050-5
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935