| Literature DB >> 12899984 |
E M P J Garrido1, J M P J Garrido, F Borges, C Delerue-Matos.
Abstract
A square-wave voltammetric (SWV) method and a flow injection analysis system with amperometric detection were developed for the determination of tramadol hydrochloride. The SWV method enables the determination of tramadol over the concentration range of 15-75 microM with a detection limit of 2.2 microM. Tramadol could be determined in concentrations between 9 and 50 microM at a sampling rate of 90 h(-1), with a detection limit of 1.7 microM using the flow injection system. The electrochemical methods developed were successfully applied to the determination of tramadol in pharmaceutical dosage forms, without any pre-treatment of the samples. Recovery trials were performed to assess the accuracy of the results; the values were between 97 and 102% for both methods.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12899984 DOI: 10.1016/s0731-7085(03)00199-7
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935