G Basileo1, Massimo Breda, G Fonte, R Pisano, C A James.
Abstract
This paper describes a high-throughput sample preparation procedure combined with LC-MS/MS analysis to measure paclitaxel in human plasma. Paclitaxel and an internal standard were extracted from plasma by a semi-automated robotic method using liquid-liquid extraction. Thereafter compounds were separated on a RP C18 column. Detection was by a PE Sciex API 3000 mass spectrometer equipped with a TurboIonSpray interface. The compounds were detected in positive ion mode using the mass transition m/z 854.6-->286.2 and m/z 831.6-->263.2 for paclitaxel and the internal standard, respectively. The limit of quantitation for paclitaxel was 1 ng/ml with an imprecision of 5.2% following extraction of 0.1 ml of plasma. Linearity was confirmed over the whole calibration range (1-1000 ng/ml) with correlation coefficients higher than 0.99 indicating good fits of the regression models. The inter and intra-day precision was better than 9.5% and the accuracy ranged from 90.3 to 104.4%. The assay was simple, fast, specific and exhibited excellent ruggedness.Entities: