Literature DB >> 12892469

Primary tissue failure of the Freestyle stentless aortic root bioprosthesis.

Takashi Kajiwara1, Yutaka Imoto, Masato Sakamoto, Kenichiro Tanaka, Takashi Matsumoto, Akira Sese.   

Abstract

We present a case of early tissue failure of the Medtronic Freestyle stentless bioprosthesis used in the aortic position which led to reoperation 27 months after implantation. Almost one half of the left coronary cusp of the prosthetic valve had torn away along the annulus. The tear started from the left-non coronary commissure and reached the midpoint of the base of the left coronary cusp. The prosthetic valve was easily excised and replaced with a new mechanical valve. There was no sign of calcification or vegetation in the excised valve. The cause of the valve dysfunction was thought to be primary tissue failure. Long-term function of the Freestyle stentless valve should be carefully followed up.

Entities:  

Mesh:

Year:  2003        PMID: 12892469     DOI: 10.1007/bf02719390

Source DB:  PubMed          Journal:  Jpn J Thorac Cardiovasc Surg        ISSN: 1344-4964


  7 in total

1.  Clinical and hemodynamic performance of the Freestyle aortic root bioprosthesis.

Authors:  P C Cartier; J G Dumesnil; J Métras; D Desaulniers; D P Doyle; M D Lemieux; G Raymond
Journal:  Ann Thorac Surg       Date:  1999-02       Impact factor: 4.330

2.  Aortic valve replacement with the freestyle stentless bioprosthesis with respect to spacial orientation of patient coronary ostia.

Authors:  A Kalangos; P Trigo-Trindade; D Vala; A Panos; B Faidutti
Journal:  J Thorac Cardiovasc Surg       Date:  2000-06       Impact factor: 5.209

3.  Outcome and follow-up of aortic valve replacement with the freestyle stentless bioprosthesis.

Authors:  A P Kappetein; J Braun; L H Baur; A Prat; K Peels; M G Hazekamp; P H Schoof; H A Huysmans
Journal:  Ann Thorac Surg       Date:  2001-02       Impact factor: 4.330

4.  Valve replacement with a stentless bioprosthesis: versatility of the porcine aortic root.

Authors:  S Westaby; X Y Jin; T Katsumata; A Arifi; P Braidley
Journal:  J Thorac Cardiovasc Surg       Date:  1998-09       Impact factor: 5.209

5.  Aortic valve replacement with the freestyle stentless bioprosthesis: five-year experience.

Authors:  K L Yun; C F Sintek; A D Fletcher; T A Pfeffer; G S Kochamba; M R Hyde; J O Torpoco; S Khonsari
Journal:  Circulation       Date:  1999-11-09       Impact factor: 29.690

6.  Time-related hemodynamic changes after aortic replacement with the freestyle stentless xenograft.

Authors:  S Westaby; N Amarasena; V Long; A Prothero; G A Amarasena; A P Banning; R Pillai; O Ormerod
Journal:  Ann Thorac Surg       Date:  1995-12       Impact factor: 4.330

7.  Aortic valve replacement with stentless porcine aortic valves: a ten-year experience.

Authors:  T E David; C M Feindel; H E Scully; J Bos; H Rakowski
Journal:  J Heart Valve Dis       Date:  1998-05
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.