Abstract

We report our initial experience with a new 18 mm Amplatzer PFO device for the closure of small PFO. Patients were selected by a semiquantitative sizing procedure using the circle of a diagnostic pigtail catheter instead of a sizing balloon. In all eight patients, the device was easy to implant without problems. Due to its size, it has the potential to minimize the risk of inadvertent atrial erosion or perforation after deployment and thus will enhance the safety of interventional PFO closure. Copyright 2003 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2003        PMID: 12891619     DOI: 10.1002/ccd.10577

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  Penetration of left and right atrial wall and aortic root by an Amplatzer atrial septal occluder in a nine year old boy with Marfan syndrome: Case report.

Authors:  Florian Loeffelbein; Christian Schlensak; Sven Dittrich
Journal:  J Cardiothorac Surg       Date:  2008-05-06       Impact factor: 1.637