Literature DB >> 12884336

Fat suppression gradient-echo magnetic resonance imaging of experimental articular cartilage lesions: comparison between phase-contrast method at 0.23T and chemical shift selective method at 1.5T.

Kari Palosaari1, Risto Ojala, Roberto Blanco-Sequeiros, Osmo Tervonen.   

Abstract

PURPOSE: To evaluate the diagnostic performance of a newly developed single-scan phase-contrast water-fat imaging technique for fat suppression at 0.23T open magnet, compared to the conventional chemical shift selective fat suppression method at 1.5T, in the detection of experimental articular cartilage lesions.
MATERIALS AND METHODS: Sixty regions of 20 knee joint specimens of pigs with artificially created articular cartilage lesions were examined with 0.23T and 1.5T MR scanners. Sagittal fat-suppressed three-dimensional gradient-echo (3D GRE) images, obtained with the phase-contrast method at 0.23T, and fat-suppressed three-dimensional spoiled gradient recalled echo (3D SPGR) images, obtained with a chemical shift selective method at 1.5T, were evaluated. Diagnostic performance was analyzed. The conspicuity of the lesions, the amount of artifacts, and the uniformity of fat suppression were evaluated. The contrast-to-noise (CNR) values of cartilage-to-bone marrow, and cartilage-to-infrapatellar fat were calculated.
RESULTS: At 0.23T, sensitivity and specificity were 80% and 95% for partial cartilage lesions (grade 2), and 91% and 100% for full-thickness lesions (grade 3). At 1.5T, sensitivity and specificity were 85% and 95% for grade 2 lesions, and 96% and 97% for grade 3 lesions. No significant difference was detected in the conspicuity of lesions. The uniformity of fat suppression was more constant with 3D SPGR images compared to 3D GRE images. More susceptibility artifacts, derived from the procedure of creating lesions, were detected at 1.5T. The cartilage-to-fat CNRs were significantly higher with high-field images.
CONCLUSION: Phase-contrast method for fat suppression at 0.23T is a useful technique in detecting articular cartilage lesions. Copyright 2003 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2003        PMID: 12884336     DOI: 10.1002/jmri.10343

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  2 in total

1.  MR-guided breast biopsy and hook wire marking using a low-field (0.23 T) scanner with optical instrument tracking.

Authors:  Roberto Blanco Sequeiros; Heli Reinikainen; Andreas M Blanco Sequeiros; Teuvo Vaara; Risto Ojala; Eija Pääkkö; Osmo Tervonen
Journal:  Eur Radiol       Date:  2006-09-22       Impact factor: 5.315

2.  MRI-guided percutaneous retrograde drilling of osteochondritis dissecans of the knee.

Authors:  Risto Ojala; Pekka Kerimaa; Martti Lakovaara; Pekka Hyvönen; Petri Lehenkari; Osmo Tervonen; Roberto Blanco-Sequeiros
Journal:  Skeletal Radiol       Date:  2011-02-15       Impact factor: 2.199

  2 in total