Literature DB >> 12883178

Design of a retractable intramedullary nail for the humerus.

A H Murdoch1, D E T Shepherd, K J Mathias, E C Stevenson.   

Abstract

This paper describes the design of a retractable intramedullary nail for the humerus that does not require inter-locking screws. The developed nail has a series of fins which open out from the casing to grip the medullary canal of the bone, thus securing it in position. Prototypes of the nail have been mechanically tested using static compression, dynamic compression and static torsion tests. During the compression tests no nails were found to fail. Statically a mean force of 809 N was withstood. During dynamic testing all the nails survived 1 million cycles, with a maximum applied load of 400 N. In torsion the two nails failed at 1.0 and 2.2 N m. Based on the mechanical testing, the retractable intramedullary nail would appear strong enough to withstand the expected loading conditions in the human body.

Entities:  

Mesh:

Year:  2003        PMID: 12883178

Source DB:  PubMed          Journal:  Biomed Mater Eng        ISSN: 0959-2989            Impact factor:   1.300


  2 in total

1.  A review of the design process for implantable orthopedic medical devices.

Authors:  G A Aitchison; D W L Hukins; J J Parry; D E T Shepherd; S G Trotman
Journal:  Open Biomed Eng J       Date:  2009-07-02

Review 2.  The Applications of Finite Element Analysis in Proximal Humeral Fractures.

Authors:  Yongyu Ye; Wei You; Weimin Zhu; Jiaming Cui; Kang Chen; Daping Wang
Journal:  Comput Math Methods Med       Date:  2017-09-10       Impact factor: 2.238

  2 in total