Literature DB >> 12880461

Comparison of the Retinomax autorefractor with hand-held retinoscopy in 1-year-old infants.

Glen A Gole1, Philip J Schluter, Julie Hall, Deborah Colville.   

Abstract

This study aimed to determine the accuracy (and usability) of the Retinomax, a hand-held autorefractor, compared to measurements taken from hand-held retinoscopy (HHR) in a sample of normal 1-year-old children. The study was a method comparison set at four Community Child Health Clinics. Infants (n = 2079) of approximately 1 year of age were identified from birth/immunization records and their caregivers were contacted by mail. A total of 327 infants ranging in age from 46 weeks to 81 weeks (mean 61 weeks) participated in the study. The children underwent a full ophthalmic examination. Under cycloplegia, refraction was measured in each eye by streak retinoscopy (HHR) and then re-measured using the Retinomax autorefractor. Sphere, cylinder, axis of cylinder and spherical equivalent measurements were recorded for HHR and Retinomax instruments, and compared. Across the range of refractive errors measured, there was generally close agreement between the two examination methods, although the Retinomax consistently read around 0.3 D less hyperopic than HHR. Significantly more girls (72 infants, 47.7%), struggled during examination with the Retinomax than boys (52 infants, 29.5%) (P < 0.001). Agreement deteriorated between the two instruments if the patient struggled during the examination (P < 0.001). In general, the Retinomax would appear to be a useful screening instrument in early childhood. However, patient cooperation affects the accuracy of results and is an important consideration in determining whether this screening instrument should be adopted for measuring refractive errors in early infancy.

Entities:  

Mesh:

Year:  2003        PMID: 12880461     DOI: 10.1046/j.1442-9071.2003.00668.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  6 in total

1.  Receding and disparity cues aid relaxation of accommodation.

Authors:  Anna M Horwood; Patricia M Riddell
Journal:  Optom Vis Sci       Date:  2009-11       Impact factor: 1.973

2.  Comparison of the Retinomax hand-held autorefractor versus table-top autorefractor and retinoscopy.

Authors:  Ibrahim Tuncer; Mehmet Ozgur Zengin; Eyyup Karahan
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

3.  Utility of an open field Shack-Hartmann aberrometer for measurement of refractive error in infants and young children.

Authors:  Erin M Harvey; Joseph M Miller; Jim Schwiegerling
Journal:  J AAPOS       Date:  2013-10       Impact factor: 1.220

4.  Refractive errors in 3-6 year-old Chinese children: a very low prevalence of myopia?

Authors:  Weizhong Lan; Feng Zhao; Lixia Lin; Zhen Li; Junwen Zeng; Zhikuan Yang; Ian G Morgan
Journal:  PLoS One       Date:  2013-10-30       Impact factor: 3.240

5.  Ellipsoid Spectacle Comparison of Plusoptix, Retinomax and 2WIN Autorefractors.

Authors:  Robert W Arnold; Samuel J Martin; Joshua R Beveridge; Andrew W Arnold; Stephanie L Arnold; Nathanael R Beveridge; Kyle A Smith
Journal:  Clin Ophthalmol       Date:  2021-08-30

6.  Ocular Refraction at Birth and Its Development During the First Year of Life in a Large Cohort of Babies in a Single Center in Northern Italy.

Authors:  Francesco Semeraro; Eliana Forbice; Giuseppe Nascimbeni; Salvatore Cillino; Vincenza Maria Elena Bonfiglio; Maria Elena Filippelli; Silvia Bartollino; Ciro Costagliola
Journal:  Front Pediatr       Date:  2020-01-29       Impact factor: 3.418