Literature DB >> 12880360

Near-infrared transmittance pulse oximetry with laser diodes.

Sonnia Maria Lopez Silva1, Maria Luisa Dotor Castilla, Juan Pedro Silveira Martin.   

Abstract

Pulse oximeters are widely used for noninvasive monitoring of oxygen saturation in arterial blood hemoglobin. We present a transmittance pulse oximetry system based on near-infrared (NIR) laser diodes (750 and 850 nm) for monitoring oxygen saturation of arterial blood hemoglobin. The pulse oximetry system is made up of the optical sensor, sensor electronics, and processing block. Also, we show experimental results obtained during the development of the whole NIR transmittance pulse oximetry system along with modifications in the sensor configuration, signal processing algorithm, and calibration procedure. Issues concerning wavelength selection and its implications for the improvement of the transmittance pulse oximetry technique are discussed. The results obtained demonstrate the proposed system's usefulness in monitoring a wide range of oxygen saturation levels. (c) 2003 Society of Photo-Optical Instrumentation Engineers.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12880360     DOI: 10.1117/1.1578495

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  2 in total

1.  Blood oxygenation and flow measurements using a single 720-nm tunable V-cavity laser.

Authors:  Yafei Feng; Haoyu Deng; Xin Chen; Jian-Jun He
Journal:  Biomed Opt Express       Date:  2017-07-03       Impact factor: 3.732

2.  Integrating Sphere Finger-Photoplethysmography: Preliminary Investigation towards Practical Non-Invasive Measurement of Blood Constituents.

Authors:  Takehiro Yamakoshi; Jihyoung Lee; Kenta Matsumura; Yasuhiro Yamakoshi; Peter Rolfe; Daiki Kiyohara; Ken-ichi Yamakoshi
Journal:  PLoS One       Date:  2015-12-04       Impact factor: 3.240

  2 in total