Pseudo-cross-stimulation due to first degree AV block.

Dimitri A Sherev1, Richard H Hongo, Nora Goldschlager.   

Abstract

During routine interrogation of a dual chamber pacemaker, atrial pacing stimulus outputs were found to coincide with surface electrogram QRS complexes raising concern for cross-stimulation. Scrutiny of the native rhythm, however, confirmed this to be "pseudo-cross-stimulation" where marked first degree AV block resulted in alignment of atrial stimulus outputs with preceding QRS complexes. Although cross-stimulation is rare, familiarity with cross-stimulation and pseudo-cross-stimulation is important because its recognition will avoid erroneous diagnoses of lead dislodgment or device malfunction.

Mesh:

Year:  2003        PMID: 12877712     DOI: 10.1046/j.1460-9592.2003.t01-1-00264.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976