| Literature DB >> 12877190 |
Dirk Löffler1, Thomas A Ternes.
Abstract
A method for the determination of gentamicin residues in hospital wastewater has been developed using kanamycin as a surrogate standard. The method consists of solid-phase extraction (SPE) and detection by ion-pair chromatography with electrospray tandem mass spectrometry (LC-ES-tandem MS). The SPE was performed on a weak cation exchanger. Filtration should be avoided in the sample preparation, otherwise a significant loss of gentamicin occurs. Chromatographic separation on a C18-column was achieved using a ternary eluent containing methanol, water and 1(-1) heptafluorobutyric acid solution. Mean relative recoveries of the analytes in hospital wastewater varied between 107 and 111%. The limit of quantification (LOQ) was 0.20 microg l(-1) in hospital wastewater. Gentamicin was found in native hospital wastewater in a concentration range between 0.4 and 7.6 microg l(-1).Entities:
Mesh:
Substances:
Year: 2003 PMID: 12877190 DOI: 10.1016/s0021-9673(03)00059-1
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759