Literature DB >> 12872090

Disposable versus reusable biopsy forceps in GI endoscopy: a cost-minimization analysis.

Céline Bourguignon1, Anne-Sabine Destrumelle, Stéphane Koch, Anne Grumblat, Pierre Carayon, Christine Chopard, Marie-Christine Woronoff-Lemsi.   

Abstract

BACKGROUND: The use of disposable biopsy forceps seems to be effective and safer than reusable biopsy forceps with respect to the risk of infection transmission. The results of cost analysis studies comparing reusable versus disposable biopsy forceps are conflicting. This study compared the cost of reusable versus disposable biopsy forceps.
METHODS: A cost-minimization analysis was carried out from the viewpoint of a hospital. Direct costs were included. The study design was retrospective. For reusable biopsy forceps, the evaluation of costs included purchase prices, cleaning (chemicals, equipment, technician time), and a fee for sterilization in a centralized facility. The cost evaluation for disposable biopsy forceps included acquisition and destruction costs. Costs were expressed in United States dollars.
RESULTS: The mean number of uses was approximately 90 per reusable forceps. The cost per use of reusable biopsy forceps was 6.84 US dollars (acquisition 3.59 US dollars, cleaning 2.28 US dollars, centralized sterilization fee 0.97 US dollars). The cost per use of disposable biopsy forceps varied from 10.72 US dollars to 15.63 US dollars. Additional cost per use of disposable biopsy forceps ranged from 3.88 US dollars to 8.79 US dollars.
CONCLUSIONS: From a strictly economic point of view, the use of reusable biopsy forceps is advantageous. However, additional factors should be considered. It would be essential to take into account the potential risk of infectious disease transmission related to the use of both types of forceps, which remains uncharacterized.

Entities:  

Mesh:

Year:  2003        PMID: 12872090     DOI: 10.1067/mge.2003.341

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  3 in total

1.  Safety-cost trade-offs in medical device reuse: a Markov decision process model.

Authors:  Thomas W Sloan
Journal:  Health Care Manag Sci       Date:  2007-02

2.  Single-use versus reusable medical devices in spinal fusion surgery: a hospital micro-costing analysis.

Authors:  C Bouthors; J Nguyen; L Durand; A Dubory; S Raspaud; Ch Court
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-07-26

3.  Performance and cost of disposable biopsy forceps in upper gastrointestinal endoscopy: comparison with reusable biopsy forceps.

Authors:  Chul-Hyun Lim; Myung-Gyu Choi; Won Chul Kim; Jin Soo Kim; Yu Kyung Cho; Jae Myung Park; In Seok Lee; Sang Woo Kim; Kyu Yong Choi; In-Sik Chung
Journal:  Clin Endosc       Date:  2012-03-31
  3 in total