Colette Fierens1, Dietmar Stöckl, Dimitri Baetens, André P De Leenheer, Linda M Thienpont.
Abstract
This study applied electrospray ionization-isotope dilution-liquid chromatography-tandem mass spectrometry for the evaluation of five urinary C-peptide immunoassays via split-sample measurements. The immunoassays measured in duplicate in the same run, the comparison method in triplicate over different runs. From the data, the within-run imprecision and the method comparison total RSDs were calculated. Regression analysis revealed on the one hand systematic differences, on the other, an excellent correlation between the test and comparison methods. From the spread of the data around the regression line in comparison with the 95% prediction intervals from the total RSD, sample-related effects and/or specificity problems were apparent and investigated.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12860032 DOI: 10.1016/s1570-0232(03)00268-x
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205