Literature DB >> 12848354

Experimental study of cryogen spray properties for application in dermatologic laser surgery.

Guillermo Aguilar1, Boris Majaron, Emil Karapetian, Enrique J Lavernia, J Stuart Nelson.   

Abstract

Cryogenic sprays are used for cooling human skin during laser dermatologic surgery. In this paper, six straight-tube nozzles are characterized by photographs of cryogenic spray shapes, as well as measurements of average droplet diameter, velocity, and temperature. A single-droplet evaporation model to predict average spray droplet diameter and temperature is tested using the experimental data presented here. The results show two distinct spray patterns--sprays for 1.4-mm-diameter nozzles (wide nozzles) show significantly larger average droplet diameters and higher temperatures as a function of distance from the nozzle compared with those for 0.5-0.8-mm-diameter nozzles (narrow nozzles). These results complement and support previously reported studies, indicating that wide nozzles induce more efficient heat extraction than the narrow nozzles.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12848354     DOI: 10.1109/TBME.2003.813537

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  1 in total

1.  Fluid and thermal dynamics of cryogen sprays impinging on a human tissue phantom.

Authors:  Walfre Franco; Henry Vu; Wangcun Jia; J Stuart Nelson; Guillermo Aguilar
Journal:  J Biomech Eng       Date:  2008-10       Impact factor: 2.097

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.