Literature DB >> 12847199

Temperature changes in perivenous tissue during endovenous laser treatment in a swine model.

Steven E Zimmet1, Robert J Min.   

Abstract

PURPOSE: To conduct a pilot study to measure temperature at the outer vein wall during endovenous laser treatment (EVLT).
METHOD: Temperature at the outer vein wall was monitored during EVLT in a live pig ear vein (8 W: 1.0 and 2.0 seconds pulse duration; 10 W: 1.0 and 1.5 second pulse duration; 12 W: 0.5, 1.0 and 1.5 second pulse duration) and exposed hind limb vein (8 W: 0.5, 1.0, 1.5 second pulse duration; 12 W: 0.5,1.0, 1.5 second pulse duration with perivenous tumescent fluid (TF); and 15 W: 0.5 second pulse duration without and with TF, 1.0 second pulse duration with TF).
RESULTS: Peak temperatures, near the outer vein wall in an ear vein of a live pig, with laser fluence at 8 W were 40.8 degrees C and 48.9 degrees C (pulse durations of 1.0 and 2.0 seconds, respectively). At 10 W, peak temperature was 47.1 degrees C and 49.1 degrees C (pulse durations of 1.0 and 1.5 seconds, respectively). At 12 W, peak temperature ranged from 37.9 degrees C (0.5 second pulse duration) to 49.1 degrees C (1.5 second pulse durations). In an exposed hind limb vessel, at 8 W, peak temperature ranged between 34.6 degrees C to 38.5 degrees C (0.5, 1.0 and 1.5 second pulse durations). At 12 W and 0.5 to 1.5 second pulse durations, with TF, peak temperature ranged from 35.6 degrees C to 39.4 degrees C. At 15 W and 0.5 second pulse duration, peak temperature was 44.0 degrees C without TF and 34.5 degrees C with TF. At 15 W and 1.0 second pulse duration, with TF, pulse duration peak temperature was 37.0 degrees C.
CONCLUSIONS: In the model studied, peak temperatures of perivenous tissues generated during endovenous laser seem unlikely to cause permanent damage to these perivenous tissues. The peak temperature generated is reduced with the use of perivenous tumescent fluid.

Entities:  

Mesh:

Year:  2003        PMID: 12847199     DOI: 10.1097/01.rvi.0000082811.60648.a4

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Optical-thermal mathematical model for endovenous laser ablation of varicose veins.

Authors:  Peter W M van Ruijven; Anna A Poluektova; Martin J C van Gemert; H A Martino Neumann; Tamar Nijsten; Cees W M van der Geld
Journal:  Lasers Med Sci       Date:  2013-10-09       Impact factor: 3.161

Review 2.  S2k guidelines: diagnosis and treatment of varicose veins.

Authors:  F Pannier; T Noppeney; J Alm; F X Breu; G Bruning; I Flessenkämper; H Gerlach; K Hartmann; B Kahle; H Kluess; E Mendoza; D Mühlberger; A Mumme; H Nüllen; K Rass; S Reich-Schupke; D Stenger; M Stücker; C G Schmedt; T Schwarz; J Tesmann; J Teßarek; S Werth; E Valesky
Journal:  Hautarzt       Date:  2022-04-19       Impact factor: 1.198

  2 in total