Investigations into the use of a reverse scan in a quadrupole ion trap mass spectrometer.

G Dobson1, J Murrell, D Despeyroux, F Wind, J-C Tabet.   

Abstract

The forward scan (i.e. an increasing RF voltage ramp for the mass-selective instability scan) is commonly used as an analytical scan for ion detection with quadrupole ion trap instruments. A number of phenomena have been observed while using this scan technique. These include space charge effects resulting in the delayed ejection of ions from the ion trap, and the fragmentation of fragile ions producing very broad peaks. Here the use of a reverse scan (i.e. a decreasing RF voltage ramp) is examined to determine the effect of the above phenomena in this acquisition method. With regard to space charge effects, the apparent reduction of the carbon isotope spacing below one Thomson (for singly charged ions) that is observed with the forward scan is now replaced by an apparent increase in this spacing. The reverse scan, which optimizes at lower axial modulation ejection voltages than the forward scan, allows for the intact ejection of fragile ions under its typical operating conditions whereas the forward scan results in fragmentation. Reducing the axial modulation voltage for the ejection of ions in the forward scan results in less dissociation of the fragile ions during ion ejection, but with the observation of ghost peaks due to incomplete ejection of all of the ions at the resonance ejection condition. While performing the reverse scan experiment, the formation of product ions from dissociation of the MH(+) ion has also been observed. Copyright 2003 John Wiley & Sons, Ltd.

Entities:  

Year:  2003        PMID: 12845593     DOI: 10.1002/rcm.1101

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  1 in total

1.  Dynamic collision-induced dissociation (DCID) in a quadrupole ion trap using a two-frequency excitation waveform: I. Effects of excitation frequency and phase angle.

Authors:  Unige A Laskay; Jennifer J Hyland; Glen P Jackson
Journal:  J Am Soc Mass Spectrom       Date:  2007-02-01       Impact factor: 3.109

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.