Ultrasound biomicroscopy and Scheimpflug photography of angle-supported phakic intraocular lens for high myopia.

Renate Ferreira de Souza1, Norma Allemann, Adriana Forseto, Paulo S Moraes Barros, Wallace Chamon, Walton Nosé.   

Abstract

PURPOSE: To evaluate the equivalence of Scheimpflug photography (SP) and ultrasound biomicroscopy (UBM) in determining corneal epithelium-intraocular lens (IOL) and border IOL-iris distances.
SETTING: Universität Erlangen-Nuremberg, Erlangen, Germany.
METHODS: In 26 eyes of 17 patients who had a NuVita MA20 angle-supported anterior chamber intraocular lens (Chiron-Domilens), SP and UBM were used to evaluate the distance between the endothelium and the anterior IOL face in central and peripheral regions (12 o'clock and 6 o'clock positions) and between the border of the anterior IOL face and the iris. The Wilcoxon test was used for statistical analysis.
RESULTS: The mean central endothelium-anterior IOL face distance was 2.01 mm and 2.00 mm by SP and UBM, respectively. The mean peripheral endothelium-anterior IOL border distance was 1.28 mm and 1.58 mm, respectively, and the mean peripheral anterior IOL face-iris distance, 0.89 mm and 0.75 mm, respectively.
CONCLUSIONS: The difference between the 2 methods in the central endothelium-anterior IOL face distance was not significant (methods were equivalent), but the difference in the peripheral endothelium-anterior IOL face distance was. This may be the result of difficulty in obtaining the exact transition point between the IOL and the haptics by SP examination. The difference between the 2 methods in the IOL border-iris distance was also significant because of the irregularity of the iris surface; therefore, measurements were performed at different sites along this structure. The significant differences in the peripheral endothelium-IOL and IOL border-iris distances indicate that although both methods are useful, they are not equivalent.

Entities:  

Mesh:

Year:  2003        PMID: 12842684     DOI: 10.1016/s0886-3350(02)02045-x

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

Review 1.  [Phakic intraocular lenses. Current status and limitations].

Authors:  H B Dick; M Tehrani
Journal:  Ophthalmologe       Date:  2004-03       Impact factor: 1.059

2.  Ultrasonography of the Anterior Segment.

Authors:  Martin Heur; Bennie H Jeng
Journal:  Ultrasound Clin       Date:  2008-04-01

Review 3.  [Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation].

Authors:  T Kohnen; T Kasper; E Terzi
Journal:  Ophthalmologe       Date:  2005-11       Impact factor: 1.059

Review 4.  [Intraocular lenses for the correction of refraction errors. Part 1: phakic anterior chamber lenses].

Authors:  T Kohnen; M Baumeister; M Cichocki
Journal:  Ophthalmologe       Date:  2005-10       Impact factor: 1.059

5.  Explaining The Current Role Of High Frequency Ultrasound In Ophthalmic Diagnosis (Ophthalmic Ultrasound).

Authors:  D Jackson Coleman; Ronald H Silverman; Mark J Rondeau; Harriet O Lloyd; Suzanne Daly
Journal:  Expert Rev Ophthalmol       Date:  2006-10-01

Review 6.  [Scheimpflug photography for the examination of phakic intraocular lenses].

Authors:  M Baumeister
Journal:  Ophthalmologe       Date:  2014-10       Impact factor: 1.059

  6 in total