| Literature DB >> 12830892 |
Masanobu Mori1, Kazuhiko Tanaka, Murad I H Helaleh, Qun Xu, Mikaru Ikedo, Yutaka Ogura, Shinji Sato, Wenzhi Hu, Kiyoshi Hasebe.
Abstract
This paper describes an ion-exclusion chromatographic system for the rapid and selective determination of ammonium ion. The optimized ion-exclusion chromatographic system was established with a polymethacrylate-based weakly basic anion-exchange resin column (TSKgel DEAE-5PW) as the separation column, an aqueous solution containing 0.05 mM tetramethylammonium hydroxide (pH 9.10) as eluent with conductimetric detection for the analyte determination. Under the optimum chromatographic conditions, ammonium ion was determined within 2.3 min with a detection limit (S/N=3) better than 0.125 microM. Ammonium ion in rain and river waters was precisely determined using this ion-exclusion chromatographic system.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12830892 DOI: 10.1016/s0021-9673(03)00625-3
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759