Crossover study of three methods for low-level white blood cell counting on two types of flow cytometer.

M J Dijkstra-Tiekstra1, J G Schrijver, P F van der Meer, R F Laport, J W Gratama, W H B M Levering, C J van Delden, J de Wildt-Eggen.   

Abstract

BACKGROUND: Flow cytometric methods were previously shown to be preferable to microscopic and volumetric methods for counting residual white blood cells (WBCs). In this study, three flow cytometric, low-level WBC counting methods were cross compared using two flow cytometers.
METHODS: Double-filtered red cell and platelet concentrates were spiked with different amounts of WBC to obtain panels of unspiked and 0.3, 1.0, 3.3, and 10.0 WBC/microl. The methods of BD Biosciences (BDB), Beckman-Coulter (BC), and an in-house method were performed on flow cytometers from BDB and BC. Samples were measured in ninefold. We required that (a) r(2) be at least 0.98 (linearity), (b) at least 80% of observations fell within 20% of expected values (accuracy), and (c) the coefficients of variation be at least 20% (precision) for samples containing at least 3.3 WBC/microl.
RESULTS: For the red cell panel, our requirements were met by the BDB method on both flow cytometers and by the BC and in-house methods on the BDB flow cytometer only. For the platelet panel, our requirements were met on all combinations of methods and flow cytometers, except for the in-house method on the BDB flow cytometer. Intra-assay variation was lowest for the BDB method, irrespective of the type of flow cytometer used.
CONCLUSION: Based on accuracy and precision, the BDB method on the BDB flow cytometer produced the best results for counting low-level WBCs. Copyright 2003 Wiley-Liss, Inc.

Mesh:

Year:  2003        PMID: 12827666     DOI: 10.1002/cyto.b.10015

Source DB:  PubMed          Journal:  Cytometry B Clin Cytom        ISSN: 1552-4949            Impact factor:   3.058


  2 in total

1.  The use of a hematology analyzer with a new generation of software as an alternative to flow cytometry for enumerating residual white blood cells in blood components.

Authors:  Richard Alejo Blanco; Chloe Cavagnetto; Laura Willmott; Elif Aydogdu; Nicola Akinyemi; Helena Standring; Simon Procter; Stephen F Garner; Atsushi Shirakami; Jarob Saker; Joachim Linssen; Rebecca Cardigan
Journal:  Transfusion       Date:  2019-11-26       Impact factor: 3.157

2.  Comparison of Flow Cytometric Methods for the Enumeration of Residual Leucocytes in Leucoreduced Blood Products: A Multicenter Study.

Authors:  Yang Zeng; Michelle Dabay; Virginia George; Shalini Seetharaman; Monika de Arruda Indig; Sharon Graminske; Nicole Kimpel; Anna Schmidt; Amanda Boerner; Sarai Paradiso; Tatyana Delman; Yunyao Li; Viktoriya Litvak; Farzad Oreizy; Angela Chen; Maryam Saleminik; Fred Mosqueda; Anna Lin; Kevin Judge
Journal:  Cytometry A       Date:  2018-01-18       Impact factor: 4.355