Preston S Wilson1, Ronald A Roy, William M Carey.
Abstract
A water-filled impedance tube capable of improved measurement accuracy and precision is reported. The measurement instrument employs a variation of the standardized two-sensor transfer function technique. Performance improvements were achieved through minimization of elastic waveguide effects and through the use of sound-hard wall-mounted acoustic pressure sensors. Acoustic propagation inside the water-filled impedance tube was found to be well described by a plane wave model, which is a necessary condition for the technique. Measurements of the impedance of a pressure-release terminated transmission line, and the reflection coefficient from a water/air interface, were used to verify the system.Entities:
Year: 2003 PMID: 12822797 DOI: 10.1121/1.1572140
Source DB: PubMed Journal: J Acoust Soc Am ISSN: 0001-4966 Impact factor: 1.840