Literature DB >> 12820741

The AbioCor implantable replacement heart.

Robert D Dowling1, Laman A Gray, Steven W Etoch, Hillel Laks, Daniel Marelli, Louis Samuels, John Entwistle, Greg Couper, Gus J Vlahakes, O H Frazier.   

Abstract

The AbioCor implantable replacement heart (IRH) is the first available totally implantable artificial heart. We recently initiated a multicenter trial of this device in patients with severe, irreversible biventricular failure. Patients who were not candidates for other therapies, including transplantation, were evaluated. All candidates were adults with inotrope-dependent biventricular failure, whose 30-day predicted mortality was higher than 70%. A three-dimensional computerized fit study predicted fit of the AbioCor thoracic unit in all recipients. At operation, the internal battery controller and transcutaneous energy transfer unit were placed. The AbioCor thoracic unit was placed in an orthotopic position after incision of the ventricals. There were 2 intraoperative deaths (due to intraoperative bleeding or aprotinin reaction). Four late deaths were recorded, 1 from multisystem organ failure and 3 cerebrovascular accidents. Autopsy revealed thrombus on the atrial struts of the 3 patients with cerebrovascular accident. Blood pumps and valves were clean on all patients. Significant morbidity was observed, primarily related to preexisting severity of illness. However, 3 patients recovered to the point of being able to take multiple trips outside of the hospital. Two patients were discharged from the hospital, with 1 patient being discharged home for more than 7 months. No significant device malfunctions or multi-system organ failure device-related infections were noted. The AbioCor IRH may be effective therapy for patients with end-stage heart failure. Many milestones have been achieved in the initial trial in humans, including the successful discharge of a patient to home and no significant device malfunctions. The occurrence of stroke is likely related to the presence of thrombus on the atrial struts and may be decreased as these atrial struts have been removed for future clinical implants.

Entities:  

Mesh:

Year:  2003        PMID: 12820741     DOI: 10.1016/s0003-4975(03)00484-3

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  10 in total

1.  Thirty-five years of mechanical circulatory support at the Texas Heart Institute: an updated overview.

Authors:  Courtney J Gemmato; Matthew D Forrester; Timothy J Myers; O H Frazier; Denton A Cooley
Journal:  Tex Heart Inst J       Date:  2005

2.  Catheterization of the AbioCor implantable replacement heart: evaluation of the unique physiology created by the device.

Authors:  Reynolds M Delgado; Mohamad Nawar; Catalin Loghin; Timothy J Myers; Igor D Gregoric; Toni Pool; Nancy Scroggins; Branislav Radovancevic; O H Frazier
Journal:  Tex Heart Inst J       Date:  2006

Review 3.  Development of mechanical circulatory support devices at the University of Tokyo.

Authors:  Yusuke Abe; Takashi Isoyama; Itsuro Saito; Shuichi Mochizuki; Minoru Ono; Hidemoto Nakagawa; Noriyuki Taniguchi; Norihiko Mitsumune; Ayaka Sugino; Mie Mitsui; Koki Takiura; Toshiya Ono; Akimasa Kouno; Tsuneo Chinzei; Shinichi Takamoto; Kou Imachi
Journal:  J Artif Organs       Date:  2007-06-20       Impact factor: 1.731

4.  The helical flow pump with a hydrodynamic levitation impeller.

Authors:  Yusuke Abe; Kohei Ishii; Takashi Isoyama; Itsuro Saito; Yusuke Inoue; Toshiya Ono; Hidemoto Nakagawa; Emiko Nakano; Kyoko Fukazawa; Kazuhiko Ishihara; Kazuyoshi Fukunaga; Minoru Ono; Kou Imachi
Journal:  J Artif Organs       Date:  2012-08-28       Impact factor: 1.731

5.  Present and future perspectives on total artificial hearts.

Authors:  Gino Gerosa; Silvia Scuri; Laura Iop; Gianluca Torregrossa
Journal:  Ann Cardiothorac Surg       Date:  2014-11

Review 6.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

7.  A glucose biofuel cell implanted in rats.

Authors:  Philippe Cinquin; Chantal Gondran; Fabien Giroud; Simon Mazabrard; Aymeric Pellissier; François Boucher; Jean-Pierre Alcaraz; Karine Gorgy; François Lenouvel; Stéphane Mathé; Paolo Porcu; Serge Cosnier
Journal:  PLoS One       Date:  2010-05-04       Impact factor: 3.240

Review 8.  The History of Durable Left Ventricular Assist Devices and Comparison of Outcomes: HeartWare, HeartMate II, HeartMate 3, and the Future of Mechanical Circulatory Support.

Authors:  Cecilia Berardi; Claudio A Bravo; Song Li; Maziar Khorsandi; Jeffrey E Keenan; Jonathan Auld; Sunny Rockom; Jennifer A Beckman; Claudius Mahr
Journal:  J Clin Med       Date:  2022-04-05       Impact factor: 4.241

9.  Image based evaluation of mediastinal constraints for the development of a pulsatile total artificial heart.

Authors:  Andreas Johann Fritschi; Marco Laumen; Sotirios Spiliopoulos; Thomas Finocchiaro; Christina Egger; Thomas Schmitz-Rode; Gero Tenderich; Reiner Koerfer; Ulrich Steinseifer
Journal:  Biomed Eng Online       Date:  2013-08-14       Impact factor: 2.819

10.  Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps.

Authors:  Kristin Unthan; Felix Gräf; Marco Laumen; Thomas Finocchiaro; Christoph Sommer; Hermann Lanmüller; Ulrich Steinseifer
Journal:  Biomed Res Int       Date:  2015-10-25       Impact factor: 3.411