Validation of a near-infrared probe for detection of thin intracranial white matter structures.

Cole A Giller1, Hanli Liu, Prem Gurnani, Sundar Victor, Umar Yazdani, Dwight C German.   

Abstract

OBJECT: The authors have developed an intracranial near-infrared (NIR) probe that analyzes the scattering of light emitted from its tip to measure the optical properties of cerebral tissue. Despite its success in distinguishing graymatter from white matter in humans during stereotactic surgery, the limits of this instrument's resolution remain unclear. In this study, the authors determined the spatial resolution of this new probe by using a rodent model supplemented with phantom measurements and computer simulation.
METHODS: A phantom consisting of Intralipid and gelatin was constructed to resemble a layer of white matter overlying a layer of gray matter. Near-infrared measurements were obtained as the probe was inserted through the gray-white matter transition. A computer simulation of NIR measurements through a gray-white matter transition was also performed using Monte Carlo techniques. The NIR probe was then used to study 19 tracks from the cortical surface through the corpus callosum in an in vivo rodent preparation. The animals were killed and histological sections through the tracks were obtained. Data from the phantom models and computer simulations showed that the NIR probe samples a volume of tissue extending 1 to 1.5 mm in front of the probe tip (this distance is termed the "lookthrough" distance). Measurements obtained from an NIR probe passing through a thin layer of white matter consisted of an initial segment of increasing values, a maximum (peak) value, and a trailing segment of decreasing values. The length of the initial segment is the lookthrough distance, the position of the peak indicates the location of the superficial white matter boundary, and the length of the trailing segment is the thickness of the layer. These considerations were confirmed in experiments with rodents. All tracks passed through the corpus callosum, which was demonstrated as a broad peak on each NIR graph. The position of the dorsal boundary of the corpus callosum and its width (based on histological measurements) correlated well with the peak of the NIR curve and its trailing segment, respectively. The initial segments correlated well with estimates of the lookthrough distance. Five of the tracks transected the smaller anterior commissure (diameter 0.2 mm), producing a narrow NIR peak at the correct depth.
CONCLUSIONS: Data in this study confirm that the NIR probe can reliably detect and measure the thickness of layers of white matter as thin as 0.2 mm. Such resolution should be adequate to detect larger structures of interest encountered during stereotactic surgery in humans.

Entities:  

Mesh:

Year:  2003        PMID: 12816278     DOI: 10.3171/jns.2003.98.6.1299

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  7 in total

Review 1.  Stereotactic implantation of deep brain stimulation electrodes: a review of technical systems, methods and emerging tools.

Authors:  Simone Hemm; Karin Wårdell
Journal:  Med Biol Eng Comput       Date:  2010-06-02       Impact factor: 2.602

2.  A laser Doppler system for monitoring cerebral microcirculation: implementation and evaluation during neurosurgery.

Authors:  Peter Rejmstad; Gustav Åkesson; Oscar Åneman; Karin Wårdell
Journal:  Med Biol Eng Comput       Date:  2015-06-24       Impact factor: 2.602

Review 3.  Deep Brain Stimulation: Emerging Tools for Simulation, Data Analysis, and Visualization.

Authors:  Karin Wårdell; Teresa Nordin; Dorian Vogel; Peter Zsigmond; Carl-Fredrik Westin; Marwan Hariz; Simone Hemm
Journal:  Front Neurosci       Date:  2022-04-11       Impact factor: 5.152

4.  Noninvasive optical inhibition with a red-shifted microbial rhodopsin.

Authors:  Amy S Chuong; Mitra L Miri; Volker Busskamp; Gillian A C Matthews; Leah C Acker; Andreas T Sørensen; Andrew Young; Nathan C Klapoetke; Mike A Henninger; Suhasa B Kodandaramaiah; Masaaki Ogawa; Shreshtha B Ramanlal; Rachel C Bandler; Brian D Allen; Craig R Forest; Brian Y Chow; Xue Han; Yingxi Lin; Kay M Tye; Botond Roska; Jessica A Cardin; Edward S Boyden
Journal:  Nat Neurosci       Date:  2014-07-06       Impact factor: 24.884

5.  Tissue-like phantoms for near-infrared fluorescence imaging system assessment and the training of surgeons.

Authors:  Alec M De Grand; Stephen J Lomnes; Deborah S Lee; Matthew Pietrzykowski; Shunsuke Ohnishi; Timothy G Morgan; Andrew Gogbashian; Rita G Laurence; John V Frangioni
Journal:  J Biomed Opt       Date:  2006 Jan-Feb       Impact factor: 3.170

6.  Determination of detection depth of optical probe in pedicle screw measurement device.

Authors:  Weitao Li; Yangyang Liu; Zhiyu Qian
Journal:  Biomed Eng Online       Date:  2014-11-01       Impact factor: 2.819

7.  Optical monitoring of cerebral microcirculation in neurointensive care.

Authors:  Peter Rejmstad; Neda Haj-Hosseini; Oscar Åneman; Karin Wårdell
Journal:  Med Biol Eng Comput       Date:  2017-12-08       Impact factor: 2.602

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.