Electrical impedance tomography to measure pulmonary perfusion: is the reproducibility high enough for clinical practice?

H J Smit1, M L Handoko, A Vonk Noordegraaf, Th J C Faes, P E Postmus, P M J M de Vries, A Boonstra.   

Abstract

A possible clinical application of electrical impedance tomography (EIT) might be to monitor changes in the pulmonary circulation, provided the reproducibility of the EIT measurement is adequate. The purpose of this study was threefold: the intra- and inter-investigator variability of repeated measurements was investigated. Three different regions of interest (ROI) were analysed to assess the optimal ROI. Twenty-four healthy subjects and six patients were included. The Sheffield applied potential tomograph (DAS-01P, IBEES, Sheffield, UK) was used. Electrodes were attached by investigator A, and duplicate EIT measurements were performed. After detachment and 45 min of rest, the protocol was repeated by another investigator B, and afterwards by the initial investigator A. Three ROIs were analysed: whole circle, 'inner half circle' and contour. The mean difference in impedance changes between observers is presented in arbitrary units (AU) +/- SD. Finally, the influence of age, body composition and sex on the EIT result was examined. For the contour ROI, the mean difference for the intra-investigator situation was -1.44 x 10(-2) +/- 18.45 x 10(-2) AU (-0.7 +/- 9.0%), and was 5.46 x 10(-2) +/- 21.66 x 10(-2) AU (2.7 +/- 10.8%) for the inter-investigator situation. The coefficient of reproducibility of the intra- and inter-investigator reproducibility varied between 0.89 and 0.97 for all ROIs (P < 0.0001). There is a relation between impedance change and age (correlation coefficient r = -0.63, P < 0.01 for contour ROI), and between impedance change and body mass index (BMI) (r = -0.53, P < 0.05). We found a significant difference in mean impedance change between groups of males and females. In conclusion, EIT results are highly reproducible when performed by the same investigator as well as by two different investigators.

Entities:  

Mesh:

Year:  2003        PMID: 12812432     DOI: 10.1088/0967-3334/24/2/359

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  7 in total

1.  Determinants of pulmonary perfusion measured by electrical impedance tomography.

Authors:  Henk J Smit; Anton Vonk Noordegraaf; J Tim Marcus; Anco Boonstra; Peter M de Vries; Pieter E Postmus
Journal:  Eur J Appl Physiol       Date:  2004-02-21       Impact factor: 3.078

2.  Imaging cardiac activity by the D-bar method for electrical impedance tomography.

Authors:  D Isaacson; J L Mueller; J C Newell; S Siltanen
Journal:  Physiol Meas       Date:  2006-04-18       Impact factor: 2.833

Review 3.  Electrical impedance tomography: the holy grail of ventilation and perfusion monitoring?

Authors:  Steffen Leonhardt; Burkhard Lachmann
Journal:  Intensive Care Med       Date:  2012-09-20       Impact factor: 17.440

4.  Regional distribution of blood volume within the preterm infant thorax during synchronised mechanical ventilation.

Authors:  Hazel R Carlisle; Ruth K Armstrong; Peter G Davis; Andreas Schibler; Inéz Frerichs; David G Tingay
Journal:  Intensive Care Med       Date:  2010-09-21       Impact factor: 17.440

Review 5.  A narrative review of electrical impedance tomography in lung diseases with flow limitation and hyperinflation: methodologies and applications.

Authors:  Ling Sang; Zhanqi Zhao; Zhimin Lin; Xiaoqing Liu; Nanshan Zhong; Yimin Li
Journal:  Ann Transl Med       Date:  2020-12

6.  Electrical impedance tomography applied to assess matching of pulmonary ventilation and perfusion in a porcine experimental model.

Authors:  Anneli Fagerberg; Ola Stenqvist; Anders Aneman
Journal:  Crit Care       Date:  2009-03-05       Impact factor: 9.097

7.  Nanoparticle-enhanced electrical impedance detection and its potential significance in image tomography.

Authors:  Ran Liu; Cuiyun Jin; Fengjuan Song; Jing Liu
Journal:  Int J Nanomedicine       Date:  2013-01-03
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.