A novel accelerometer tilt switch device for switch actuation in the patient with profound disability.

Stephen Perring1, Anthony Summers, Emma L Jones, Frances J Bowen, Katharine Hart.   

Abstract

We describe the design and operation of a new switch that can be operated by patients with severely limited movement. The basis for the switch is an inexpensive single-chip accelerometer device. The switch responds to a relatively rapid rotation of the active components. We have ensured that the sensitivity of the switch device is adjustable over a wide range. The device automatically adjusts for changes in attitude of the device that result from a patient changing posture. We show that the device can operate in a wide range of attitudes. We describe 2 case studies in which the switch was successfully used as a head tilt switch.

Entities:  

Mesh:

Year:  2003        PMID: 12808551     DOI: 10.1016/s0003-9993(03)00013-3

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  1 in total

1.  The design and testing of a novel mechanomyogram-driven switch controlled by small eyebrow movements.

Authors:  Natasha Alves; Tom Chau
Journal:  J Neuroeng Rehabil       Date:  2010-05-21       Impact factor: 4.262