Literature DB >> 12808406

Vertical cup-to-disc ratio: agreement between direct ophthalmoscopic estimation, fundus biomicroscopic estimation, and scanning laser ophthalmoscopic measurement.

Russell Watkins1, Leena Panchal, Jimmy Uddin, Pinakin Gunvant.   

Abstract

PURPOSE: Ophthalmoscopic estimation of the vertical cup-to-disc ratio (VCDR) of the optic nerve head is important in the management of patients with glaucoma or who are glaucoma suspects. The purpose of this study was to compare the accuracy of estimation of VCDR obtained with direct ophthalmoscopy with that obtained with fundus biomicroscopy through undilated pupils. Measurements of VCDR obtained with the Heidelberg Retina Tomograph (HRT), a confocal scanning laser ophthalmoscope, were used as a standard.
METHODS: Thirty young, healthy adults had their optic discs photographed and then imaged and analyzed with the HRT. Due to HRT software limitations, the VCDR could not be calculated automatically, and so a validated VCDR measurement was derived. This was used as the standard against which ophthalmoscopic estimations were compared. Two months later, the subjects had their VCDRs estimated using direct ophthalmoscopy and fundus biomicroscopy performed in random order after varying time intervals. Agreement between ophthalmoscopic VCDR estimation and HRT-derived VCDR measurement was assessed by means of bias plots and the weighted kappa statistic.
RESULTS: Compared with the HRT-derived VCDR measurement, both ophthalmoscopic techniques tended to underestimate VCDR. The bias with direct ophthalmoscopy was statistically significant. There were also wide variations in VCDR estimation with direct ophthalmoscopy and fundus biomicroscopy, which were more pronounced with direct ophthalmoscopy. The weighted kappa statistic indicated moderate agreement between fundus biomicroscopy and HRT-derived VCDR measurement. There was poor to fair agreement between direct ophthalmoscopy and HRT-derived VCDR measurement. The level of disagreement was independent of HRT-derived VCDR, optic nerve head size and pupil size for both direct ophthalmoscopy and fundus biomicroscopy.
CONCLUSIONS: Fundus biomicroscopy enables more accurate, less variable VCDR estimation than direct ophthalmoscopy. The clinician should record which method they used to examine the optic nerve head so that subsequent clinical decisions are not influenced by apparent VCDR changes.

Entities:  

Mesh:

Year:  2003        PMID: 12808406     DOI: 10.1097/00006324-200306000-00012

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  7 in total

1.  A comparison of cup-to-disc ratio estimates by fundus biomicroscopy and stereoscopic optic disc photography in the Tema Eye Survey.

Authors:  J C Mwanza; D S Grover; D L Budenz; L W Herndon; W Nolan; J Whiteside-de Vos; G Hay-Smith; J R Bandi; K A Bhansali; L A Forbes; W J Feuer; K Barton
Journal:  Eye (Lond)       Date:  2017-04-07       Impact factor: 3.775

2.  Optic disc analysis with Heidelberg Retina Tomography III in glaucoma with unilateral visual field defects.

Authors:  Ge-Ge Xiao; Ling-Ling Wu
Journal:  Jpn J Ophthalmol       Date:  2010-08-11       Impact factor: 2.447

Review 3.  Optic disk size and glaucoma.

Authors:  Esther M Hoffmann; Linda M Zangwill; Jonathan G Crowston; Robert N Weinreb
Journal:  Surv Ophthalmol       Date:  2007 Jan-Feb       Impact factor: 6.197

4.  Does access to a portable ophthalmoscope improve skill acquisition in direct ophthalmoscopy? A method comparison study in undergraduate medical education.

Authors:  J A Gilmour-White; A Picton; A Blaikie; A K Denniston; R Blanch; J Coleman; P I Murray
Journal:  BMC Med Educ       Date:  2019-06-13       Impact factor: 2.463

5.  Comparison of the clinical estimation of cup-to-disk ratio by direct ophthalmoscopy and optical coherence tomography.

Authors:  Angela Ofeibea Amedo; Nana Yaa Koomson; Emmanuel Kobia Acquah; Tchiakpe Michel Pascal; Johnson Atuahene; Prince Kwaku Akowuah; Philip Tetteh Djeagbo; Richard Baafi
Journal:  Ther Adv Ophthalmol       Date:  2019-04-02

6.  Detecting glaucoma from multi-modal data using probabilistic deep learning.

Authors:  Xiaoqin Huang; Jian Sun; Krati Gupta; Giovanni Montesano; David P Crabb; David F Garway-Heath; Paolo Brusini; Paolo Lanzetta; Francesco Oddone; Andrew Turpin; Allison M McKendrick; Chris A Johnson; Siamak Yousefi
Journal:  Front Med (Lausanne)       Date:  2022-09-29

7.  Clinical measurement and categorization of optic disc in glaucoma patients.

Authors:  Harsha B Laxmana Rao; G C Sekhar; Ganesh J Babu; Rajul S Parikh
Journal:  Indian J Ophthalmol       Date:  2009 Sep-Oct       Impact factor: 1.848

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.