Literature DB >> 12808180

Clinical and Doppler ultrasonography data of a polyurethane vascular access graft for haemodialysis: a prospective study.

Patrick Wiese1, Jochen Blume, Hans-Joachim Mueller, Hermann Renner, And Barbara Nonnast-Daniel.   

Abstract

BACKGROUND: Absence of a permanent vascular access in most patients starting haemodialysis remains a cause of high morbidity and costs. This study obtained new clinical and colour Doppler ultrasound (CDU) data of a polyurethane vascular access graft (PVAG) proposing early post-operative cannulation.
METHODS: Baseline characteristics were determined in 15 patients and the PVAGs were evaluated prospectively including first cannulation, patency and complications. CDU was used post-operatively and after 1 year for assessing graft morphology and access blood flow.
RESULTS: PVAGs were cannulated at a median of 4 days post-operatively. The 1-year primary patency of the PVAG was 66.7%. During the 15 months observation three grafts thrombosed, one was replaced because of infection and one because of ischaemia. CDU measurements at the feeding brachial artery revealed a mean initial access volume flow of 773+/-89 ml/min, being significantly higher in patients without thrombosis compared to patients with thrombotic events (930+/-90 vs 375+/-143 ml/min, P<0.05). The initial inability to directly monitor PVAGs by CDU changed at sites of frequent centesis, where Doppler signals and luminal morphology could be evaluated in the follow up examination.
CONCLUSIONS: The PVAG offers early access for urgent haemodialysis. CDU for access volume flow measurement at the feeding brachial artery contributes to predict access thrombosis. Direct non-invasive graft imaging is limited and the ultrasonographical changes in the polyurethane material enabling graft monitoring after repeated cannulation might indicate an injury of the graft with increased risk for access failure.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12808180     DOI: 10.1093/ndt/gfg168

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  2 in total

1.  Surveillance of hemodialysis vascular access.

Authors:  William L Whittier
Journal:  Semin Intervent Radiol       Date:  2009-06       Impact factor: 1.513

2.  Patency rate and complications of polytetrafluoroethylene grafts compared with polyurethane grafts for hemodialysis access.

Authors:  Hassan Ravari; Gholam Hossein Kazemzade; Mohammad Hadi Saied Modaghegh; Patricia Khashayar
Journal:  Ups J Med Sci       Date:  2010-03-11       Impact factor: 2.384