Abstract
As a means of decontaminating instruments possibly exposed to Creutzfeldt-Jakob disease, the World Health Organization has recommended immersion and autoclaving in sodium hydroxide. However, this recommendation has raised concerns of possible damage to autoclaves, and hazards to operators as a result of the caustic vapors. A series of experiments has been conducted that demonstrate that there are containment pan-and-lid combinations in which instruments can be autoclaved in sodium hydroxide without risk to the autoclave or the operator.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12806365 DOI: 10.1067/mic.2003.51
Source DB: PubMed Journal: Am J Infect Control ISSN: 0196-6553 Impact factor: 2.918