Literature DB >> 12806050

A new all-season passive sampling system for monitoring H2S in air.

Hongmao Tang1, James Sandeluk, Linda Lin, J William Lown.   

Abstract

A new Maxxam All-Season Passive Sampling system for monitoring H2S in air has been developed. This passive sampling system employs the same approaches as the Maxxam All-Season Passive Sampling Systems for monitoring SO2, NO2, and O3 reported previously. This system has been extensively tested in the lab (temperature from -20 to 20 degrees C, relative humidity from 30 to 84%, and wind speed from 0.5 to 150 cm/s) and validated in field studies. Comparing measurements obtained with the use of the new passive sampling system with equivalent measurement with the use of an active filter pack H2S sampler yielded an accuracy of greater than 85%. The new H2S passive sampling system can be used to measure ambient H2S concentrations ranging from 0.02 to 7 ppb based on a 1-month exposure period. There is no significant interference found from other sulfur compounds in air. This system has been used in many air monitoring projects.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12806050      PMCID: PMC6009708          DOI: 10.1100/tsw.2002.87

Source DB:  PubMed          Journal:  ScientificWorldJournal        ISSN: 1537-744X


  1 in total

1.  Relative performance of different exposure modeling approaches for sulfur dioxide concentrations in the air in rural western Canada.

Authors:  Igor Burstyn; Nicola M Cherry; Yutaka Yasui; Hyang-Mi Kim
Journal:  BMC Med Res Methodol       Date:  2008-07-04       Impact factor: 4.615

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.