Resistive load of laryngeal mask airway and ProSeal laryngeal mask airway in mechanically ventilated patients.

G Natalini1, A Rosano, G Lanza, E Martinelli, C Pletti, A Bernardini.   

Abstract

BACKGROUND: The ProSeal Laryngeal Mask Airway (PLMA) ventilation tube is narrower and shorter than the standard Laryngeal Mask Airway (LMA) and is without the vertical bars at the end of the tube. In this randomized, crossover study, PLMA and LMA resistances were compared.
METHODS: Respiratory mechanics was calculated in 26 anesthetized, mechanically ventilated patients with both LMA and PLMA. The laryngeal mask positioning was fiberoptically evaluated. Differences in the respiratory mechanics of the LMA and the PLMA were attributed to the differences between the laryngeal masks.
RESULTS: In the total study population the airway resistance was 1.5 +/- 2.6 hPa.l-1.s-1 (P = 0.005) higher with the PLMA than with the LMA. During the PLMA use, the peak expiratory flow reduced by 0.02 +/- 0.05 l min-1 (P = 0.046), the expiratory resistance increased by 0.6 +/- 1.3 hPa.l-1.s-1 (P = 0.022), and the time constant of respiratory system lengthened by 0.09 +/- 0.18 s (P = 0.023). These differences doubled when the LMA was better positioned than the PLMA, whereas they disappeared when the PLMA was positioned better than the LMA.
CONCLUSIONS: The standard LMA offers a lower resistive load than the PLMA. Moreover, the fitting between the laryngeal masks and the larynx, as fiberoptically evaluated, plays a major role in determining the resistive properties of these devices.

Entities:  

Mesh:

Year:  2003        PMID: 12803596     DOI: 10.1034/j.1399-6576.2003.00133.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  5 in total

Review 1.  ProSeal versus Classic laryngeal mask airway (LMA) for positive pressure ventilation in adults undergoing elective surgery.

Authors:  Muhammad Qamarul Hoda; Khalid Samad; Hameed Ullah
Journal:  Cochrane Database Syst Rev       Date:  2017-07-20

Review 2.  Laryngeal Masks in Neonatal Resuscitation-A Narrative Review of Updates 2022.

Authors:  Srinivasan Mani; Joaquim M B Pinheiro; Munmun Rawat
Journal:  Children (Basel)       Date:  2022-05-17

3.  PLMA vs. I-gel: A Comparative Evaluation of Respiratory Mechanics in Laparoscopic Cholecystectomy.

Authors:  Bimla Sharma; Raminder Sehgal; Chand Sahai; Jayashree Sood
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2010-10

4.  Relationship Between Respiratory Dynamics and Body Mass Index in Patients Undergoing General Anesthesia with Laryngeal Mask Airway (LMA) and Comparison Between Lithotomy and Supine Positions.

Authors:  Xiao Zhao; Shiwei Huang; Zhaomin Wang; Lianhua Chen; Shitong Li
Journal:  Med Sci Monit       Date:  2016-08-01

5.  Comparison of ventilatory efficacy and airway dynamics between ProSeal laryngeal mask airway and endotracheal tube in adult patients during general anesthesia.

Authors:  Sudheesh Kannan; S S Harsoor; L Sowmiya; S S Nethra; D DevikaRani; M Sathesha
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2015 Oct-Dec
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.